Galvanic corrosion behavior of implant suprastructure dental alloys.

Nadia M Taher1, Abed S Al Jabab.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate and compare in vitro, the galvanic corrosion behavior of Co-Cr alloys (R2000, R800), Ni-Cr (RCS), silver-palladium (Jelstar), Gold (Pontallor-4) and Ternary Ti (experimental Ter Ti) when coupled with endosseous Ti implant abutment material. Amalgam alloy and commercially pure Ti cylinders (SSTi) were coupled with endosseous Ti implants as negative and positive controls, respectively.
METHODS: An EG&G Model 263 Scanning Potentiostat was used for this purpose. Specimens were prepared and fresh artificial saliva was used as an electrolyte solution. The experiment run time was 24h for each couple. The common potential, galvanic current and current integration during the last 6h were recorded for each couple.
RESULTS: The results showed that the best couples were Ti/Pontallor-4, Ti/Ter Ti, Ti/R800 and Ti/Jelstar. The least acceptable couples were Ti/amalgam, SSTi/SSTi and Ti/R2000, while the Ti/RCS couple showed unstable galvanic corrosion behavior. SIGNIFICANCE: It is concluded that the following alloys can be used as suprastructure alloys with Ti implants: Pontallor-4, R800, Jelstar and Ter Ti. Although Ter Ti alloy is an experimental alloy, it showed good results, but cannot be used in the clinical field unless extensive investigations are carried out. The SSTi/SSTi couple showed unexpected galvanic corrosion behavior which needs further investigation.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12498897     DOI: 10.1016/s0109-5641(02)00008-8

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  6 in total

1.  Release of elements from retrieved maxillofacial plates and screws.

Authors:  C Bertoldi; J M Pradelli; U Consolo; D Zaffe
Journal:  J Mater Sci Mater Med       Date:  2005-09       Impact factor: 3.896

2.  Assessment of basal implants in compromised ridges.

Authors:  M Anuradha; Harsha V Babaji; Neel V Hiremath; V A Usha; Arunoday Kumar; Tanya Nandkeoliar; Sankalp Verma
Journal:  J Family Med Prim Care       Date:  2020-04-30

Review 3.  Corrosion degradation and prevention by surface modification of biometallic materials.

Authors:  Raghuvir Singh; Narendra B Dahotre
Journal:  J Mater Sci Mater Med       Date:  2006-12-02       Impact factor: 4.727

4.  Electrochemical Behaviour and Galvanic Effects of Titanium Implants Coupled to Metallic Suprastructures in Artificial Saliva.

Authors:  Ana Mellado-Valero; Anna Igual Muñoz; Virginia Guiñón Pina; Ma Fernanda Sola-Ruiz
Journal:  Materials (Basel)       Date:  2018-01-22       Impact factor: 3.623

5.  Corrosive Studies of a Prosthetic Ni-Cr Alloy Coated with Ti(C,N) Type Layers.

Authors:  Katarzyna Banaszek; Marek Maślanka; Michael Semenov; Leszek Klimek
Journal:  Materials (Basel)       Date:  2022-03-27       Impact factor: 3.623

Review 6.  Electrogalvanism in Oral Implantology: A Systematic Review.

Authors:  Meriem Amine; Wiam Merdma; Khalid El Boussiri
Journal:  Int J Dent       Date:  2022-08-05
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.