Ability of calibration phantom to reduce the interscan variability in electron beam computed tomography.

Matthew J Budoff1, Songshou Mao, Bin Lu, Junichiro Takasu, Janis Child, Sivi Carson, Hans Fisher.   

Abstract

OBJECTIVE: To test the hypothesis that a calibration phantom would improve interpatient and interscan variability in coronary artery calcium (CAC) studies.
METHODS: We scanned 144 patients twice with or without the calibration phantom and then scanned 93 patients with a single calcific lesion twice and, finally, scanned a cork heart with calcific foci.
RESULTS: There were no linear correlations in computed tomography Hounsfield unit (CT HU) and CT HU interscan variation between blood pool and phantom plugs at any slice level in patient groups (p > 0.05). The CT HU interscan variation in phantom plugs (2.11 HU) was less than that of the blood pool (3.47 HU; p < 0.05) and CAC lesion (20.39; p < 0.001). Comparing images with and without a calibration phantom, there was a significant decrease in CT HU as well as an increase in noise and peak values in patient studies and the cork phantom study.
CONCLUSION: The CT HU attenuation variations of the interpatient and interscan blood pool, calibration phantom plug, and cork coronary arteries were not parallel. Therefore, the ability to adjust the CT HU variation of calcific lesions by a calibration phantom is problematic and may worsen the problem.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12488730     DOI: 10.1097/00004728-200211000-00005

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  2 in total

1.  Detection of small vessels with electron beam computed tomographic angiography using 1.5 and 3 mm collimator protocols.

Authors:  Songshou Mao; Jerold S Shinbane; Ronald J Oudiz; Ferdinand Flores; Alex Chau; Stephen Liu; Shaojun Wang; Yanlin Gao; Matthew J Budoff
Journal:  Int J Cardiovasc Imaging       Date:  2005-07-23       Impact factor: 2.357

Review 2.  CT for imaging coronary artery disease: defining the paradigm for its application.

Authors:  Bernd M Ohnesorge; Lars K Hofmann; Thomas G Flohr; U Joseph Schoepf
Journal:  Int J Cardiovasc Imaging       Date:  2005-02       Impact factor: 2.357

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.