[Narcotrend EEG monitoring during total intravenous anaesthesia in 4.630 patients].

W Wilhelm1, S Kreuer, R Larsen.   

Abstract

INTRODUCTION: The Narcotrend is a new EEG monitor designed to measure the hypnotic component of anaesthesia; however, a major clinical evaluation is still missing. This prospective multicentre study was designed to investigate the feasibility of Narcotrend monitoring in a large number of patients under different clinical conditions and to define its impact on recovery times after propofol-based total intravenous anaesthesia.
METHODS: After legal authority approval and patients'informed consent had been obtained, total intravenous anaesthesia was induced and maintained with propofol and an opioid analgesic at the discretion of the attending anaesthesiologist. In the first 10-15 patients of each centre the anaesthesiologist was blinded to the Narcotrend recordings and propofol was dosed according to clinical needs. In the following patients propofol was infused at a rate sufficient to achieve a target Narcotrend stage of D or E. With termination of propofol infusion,recovery times were recorded and analysed for the patients with or without Narcotrend monitoring; in addition, recovery times were analysed depending on the Narcotrend stage at the moment of termination of propofol infusion.
RESULTS: In total, 4,630 adult patients were studied at 46 institutions, 521 without and 4,109 with Narcotrend monitoring. Demographic data and duration of anaesthesia were comparable. Emergence from anaesthesia was significantly shorter in Narcotrend monitored patients, e.g.opening eyes after 9.8+/-5.9 (mean+/-SD) vs.11.8+/-7.1 min. In addition,awakening was significantly more rapid when the propofol infusion was stopped at a lower level of hypnosis as indicated by Narcotrend monitoring, e.g.opening eyes after 7.1+/-4.5 min with stage C instead of 17.0+/-7.4 min with stage F.
CONCLUSIONS: The EEG monitor Narcotrend can be used for adult patients of different ages and during various surgical procedures.Narcotrend monitoring facilitates a reduction of recovery times after propofol-based total intravenous anaesthesia,presumably by allowing for an individual titration of the propofol dosage. Moreover, it appears that the profile of recovery can be optimised when at the end of surgery,the propofol infusion is controlled to Narcotrend stage C instead of D, E, or F.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12486586     DOI: 10.1007/s00101-002-0417-3

Source DB:  PubMed          Journal:  Anaesthesist        ISSN: 0003-2417            Impact factor:   1.041


  8 in total

1.  Correlation of Narcotrend Index, entropy measures, and spectral parameters with calculated propofol effect-site concentrations during induction of propofol-remifentanil anaesthesia.

Authors:  Ulrich Grouven; Frank A Beger; Barbara Schultz; Arthur Schultz
Journal:  J Clin Monit Comput       Date:  2004-08       Impact factor: 2.502

Review 2.  [Measurement of the depth of anaesthesia].

Authors:  G N Schmidt; J Müller; P Bischoff
Journal:  Anaesthesist       Date:  2008-01       Impact factor: 1.041

3.  [Depth of anaesthesia during intubation: comparison between propofol and thiopentone].

Authors:  C E Beck; B Pohl; M Janda; J Bajorat; R Hofmockel
Journal:  Anaesthesist       Date:  2006-04       Impact factor: 1.041

4.  Comparison of Narcotrend Index, Bispectral Index, spectral and entropy parameters during induction of propofol-remifentanil anaesthesia.

Authors:  Arthur Schultz; Markus Siedenberg; Ulrich Grouven; Thomas Kneif; Barbara Schultz
Journal:  J Clin Monit Comput       Date:  2008-02-21       Impact factor: 2.502

5.  Failure of two commercial indexes and spectral parameters to reflect the pharmacodynamic effect of desflurane on EEG.

Authors:  Sascha Kreuer; Jörgen Bruhn; Richard Ellerkmann; Stephan Ziegeler; Darius Kubulus; Wolfram Wilhelm
Journal:  J Clin Monit Comput       Date:  2008-03-27       Impact factor: 2.502

6.  Clinical study of etomidate emulsion combined with remifentanil in general anesthesia.

Authors:  Digui Weng; Menghua Huang; Runnian Jiang; Rufu Zhan; Chunni Yang
Journal:  Drug Des Devel Ther       Date:  2013-08-20       Impact factor: 4.162

7.  Deep sedation for endoscopic retrograde cholangiopancreatography: a comparison between clinical assessment and Narcotrend(TM) monitoring.

Authors:  Somchai Amornyotin; Wiyada Chalayonnawin; Siriporn Kongphlay
Journal:  Med Devices (Auckl)       Date:  2011-03-17

8.  Effects of isoflurane, remifentanil and dexmedetomidine on selected EEG parameters derived from a Narcotrend Monitor before and after nociceptive stimulation at different MAC multiples in cats.

Authors:  Jonathan F Raue; Julia Tünsmeyer; Sabine B R Kästner
Journal:  BMC Vet Res       Date:  2020-09-14       Impact factor: 2.741

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.