Literature DB >> 12468129

The computer simulation of wear behavior appearing in total hip prosthesis.

James Shih-Shyn Wu1, Jui-Pin Hung, Chi-Shiang Shu, Jian-Horng Chen.   

Abstract

Computer algorithms are proposed for the estimation of wear appearing in artificial hip joints using finite element analysis based on the modified Archard's wear law, contact features and an analogue wear process. A pin-on-disk plate experiment is reconstructed to assess the efficiency and validity of the algorithms proposed here. Through the successful verification of wear depth and volume loss of the pin-on-disk plate as well as the artificial hip joint, the current algorithms provide significant agreement with experiments, clinical measurements and numerical calculations and are shown to be both valid and feasible. Further investigation into the effect of femoral heads with various sizes suggests that the larger femoral head may induce larger wear volume but gives a smaller wear depth and that wear depth and volume loss are apparently nonlinearly related to the femoral head diameter. It is shown that the current algorithms are useful and helpful in understanding wear behavior for alternative or new designs of artificial hip joints and even for other analogous structures.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12468129     DOI: 10.1016/s0169-2607(01)00199-7

Source DB:  PubMed          Journal:  Comput Methods Programs Biomed        ISSN: 0169-2607            Impact factor:   5.428


  6 in total

1.  Wear analysis of chamfered elongated acetabular cup liners.

Authors:  Hsiao-Che Lin; Tzuo-Liang Luo; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2011-12-22       Impact factor: 2.602

2.  Wear patterns of, and wear volume formulae for, cylindrically elongated acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2010-04-28       Impact factor: 2.602

3.  Using nonlinear finite element models to analyse stress distribution during subluxation and torque required for dislocation of newly developed total hip structure after prosthetic impingement.

Authors:  Wei-Min Chi; Chien-Chung Lin; Ying-Jui Ho; Hsiao-Che Lin; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2017-07-01       Impact factor: 2.602

4.  Cross-shear implementation in sliding-distance-coupled finite element analysis of wear in metal-on-polyethylene total joint arthroplasty: intervertebral total disc replacement as an illustrative application.

Authors:  Curtis M Goreham-Voss; Philip J Hyde; Richard M Hall; John Fisher; Thomas D Brown
Journal:  J Biomech       Date:  2010-06-18       Impact factor: 2.712

5.  In vitro evaluation of stiffness graded artificial hip joint femur head in terms of joint stresses distributions and dimensions: finite element study.

Authors:  H Fouad
Journal:  J Mater Sci Mater Med       Date:  2011-04-20       Impact factor: 3.896

6.  Comparative Evaluation of Carbon Reinforced Polyetherketone Acetabular Cup using Finite Element Analysis.

Authors:  Abdal A; Noorani R; Cha G
Journal:  J Biomed Phys Eng       Date:  2020-12-01
  6 in total