Peter C Wever1, Yvonne M C Henskens, Piet A Kager, Jacob Dankert, Tom van Gool.
Abstract
The sensitivity and specificity of the Cell-Dyn 4000 hematology analyzer in the diagnosis of imported malaria were studied with samples from patients in an academic hospital setting. The performance of the Cell-Dyn 4000 hematology analyzer was compared with that of conventional diagnostic methods for malaria. The Cell-Dyn 4000 hematology analyzer detected hemozoin-containing depolarizing monocytes in 29 of 58 patients with malaria and 2 of 55 patients without malaria. The presence or absence of depolarizing monocytes in patients with malaria was related to duration of symptoms before presentation for malaria analysis. A second parameter, pseudoreticulocytosis due to nuclear material of intraerythrocytic malaria parasites, was detected by the Cell-Dyn 4000 hematology analyzer almost exclusively in Plasmodium falciparum malaria patients with parasitemia levels of >/=0.5%. Attention to these abnormalities in medical centers without tropical disease expertise may decrease a delay in the diagnosis of malaria.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12454179 PMCID: PMC154588 DOI: 10.1128/JCM.40.12.4729-4731.2002