BACKGROUND: The potentially accommodative intraocular lens (IOL) is a new development in IOL design METHODS: We evaluated the new Humanoptics 1CU accommodative IOL in a laboratory study with human post mortem autopsy eyes. Using the Miyake-Apple posterior view video technique, the movement pattern of the IOL was tested and observed from the posterior perspective. RESULTS. A circular bend at the level of the ciliary body applied slight circular force onto the sclera allowing the relaxation of the zonules. The shift of focus was demonstrated by using a reading target. In addition, viscoelastic was injected into the vitreous resulting in the same anterior movement of the IOL optic. CONCLUSIONS: The 1CU Humanoptics accommodative IOL showed potential accommodative behaviour in the laboratory. The accommodative (respectively pseudoaccommodative) effect was based on the anterior shift principle with anterior movement of the IOL-optic in the state of relaxing zonules. Whether this reflects the clinical situation, especially to this extent, must be further evaluated.
BACKGROUND: The potentially accommodative intraocular lens (IOL) is a new development in IOL design METHODS: We evaluated the new Humanoptics 1CU accommodative IOL in a laboratory study with human post mortem autopsy eyes. Using the Miyake-Apple posterior view video technique, the movement pattern of the IOL was tested and observed from the posterior perspective. RESULTS. A circular bend at the level of the ciliary body applied slight circular force onto the sclera allowing the relaxation of the zonules. The shift of focus was demonstrated by using a reading target. In addition, viscoelastic was injected into the vitreous resulting in the same anterior movement of the IOL optic. CONCLUSIONS: The 1CU Humanoptics accommodative IOL showed potential accommodative behaviour in the laboratory. The accommodative (respectively pseudoaccommodative) effect was based on the anterior shift principle with anterior movement of the IOL-optic in the state of relaxing zonules. Whether this reflects the clinical situation, especially to this extent, must be further evaluated.
Authors: J S Wolffsohn; S A Naroo; N K Motwani; S Shah; O A Hunt; S Mantry; M Sira; I A Cunliffe; M T Benson Journal: Br J Ophthalmol Date: 2006-03-10 Impact factor: 4.638