Failure of suture material at suture anchor eyelets.

Dominik C Meyer1, Richard W Nyffeler, Sandro F Fucentese, Christian Gerber.   

Abstract

PURPOSE: In the repair of soft tissue to bone using suture anchors, failure of the suture material can occur at the anchor eyelet. This study examines the load strength at which suture material fails with different metallic suture anchor eyelets. TYPE OF STUDY: Biomechanical study.
METHODS: Suture material (Ethibond No. 2, Ethicon, Norderstedt, Germany) was pulled out from 22 metallic suture anchor models at 60 mm/min, and tensile load at failure and failure mode were recorded. Tests were performed either by simultaneous pulling on 2 suture limbs in 3 different directions (straight, at 45 degrees, and at 45 degrees rotated by 90 degrees to the suture anchor axis) or by pulling on 1 suture limb while measuring the resulting force on the second limb. All tests were performed until suture failure. Pulling was performed in single tests on an Instron materials testing machine (High Wycombe, UK), with the anchors held by a vise.
RESULTS: In all cases, the suture failed at the anchor eyelet. Failure load at straight loading ranged from 116 +/- 5 N to 226 +/- 5 N and from 69 +/- 5 N to 193 +/- 7 N when loaded at an angle of 45 degrees. The best results were found with the Statak 5.2-mm (Zimmer, Warsaw, IN): 177 N; Corkscrew 6.5-mm anchor (Arthrex, Naples, FL): 174 N; and PeBA 4.0-mm anchor (OBL Orthopaedic Biosystems, Scottsdale, AZ): 169 N. With each eyelet, sutures failed preferentially in 1 direction, depending on the presence of sharp edges.
CONCLUSIONS: Suture material can be cut at suture anchor eyelets. Failure load depends on sharp edges on the eyelet and occurs at forces up to 73% below the breaking strength of the suture material on a smooth hook. Anchors with suture-protecting channels are particularly sensitive to the orientation in which the sutures are loaded.

Mesh:

Substances:

Year:  2002        PMID: 12426545     DOI: 10.1053/jars.2002.36115

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  4 in total

1.  The effect of two nonresorbable suture types on the mechanical performance over a metal suture anchor eyelet.

Authors:  D Acton; A Perry; R Evans; A Butler; P Stephens; W Bruce; J Goldberg; D Sonnabend; W R Walsh
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-09-12       Impact factor: 4.342

Review 2.  [Arthroscopic suturing of the rotator cuff. Placing of anchor, suturing and tying techniques].

Authors:  M Kettler; E Kurtoglu; J Grifka; M Tingart
Journal:  Orthopade       Date:  2007-09       Impact factor: 1.087

3.  Influence of Loading Conditions on the Mechanical Performance of Multifilament Coreless UHMWPE Sutures Used in Orthopaedic Surgery.

Authors:  Maria Prado-Novoa; Laura Perez-Sanchez; Belen Estebanez; Salvador Moreno-Vegas; Ana Perez-Blanca
Journal:  Materials (Basel)       Date:  2022-03-31       Impact factor: 3.623

4.  Biomechanical evaluation of the long head of the biceps brachii tendon fixed by three techniques: a sheep model.

Authors:  Carlos Henrique Ramos; Júlio Cezar Uili Coelho
Journal:  Rev Bras Ortop       Date:  2016-12-30
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.