This report describes a new sample preparation method for food which allows a complete separation of carbohydrates and amino acids prior to their analysis by anion-exchange chromatography and integrated pulsed amperometric detection. Food samples with high carbohydrate concentrations are applied to solid-phase extraction columns containing a strong cation-exchange resin. Carbohydrates are recovered initially; retained amino acids are eluted with 0.2 M CaC l(2) subsequently. The carbohydrate and the amino acid fractions are analyzed. The recovery calculated for 21 amino acids was in the range from 84 to 126%. The sample preparation was tested for amino acid concentrations between 4.2 and 84.0 nmol of each amino acid (between 2.1 and 42.0 nmol of cystine) and correlation coefficients between 0.84 and 0.99 were obtained. The capacity of the solid-phase extraction columns employed was up to 3.7 micro mol. Sample preparation was evaluated with four different food samples: sourdough, skim milk, lemon juice, and potato.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12423635     DOI: 10.1016/s0003-2697(02)00283-x

Source DB:  PubMed          Journal:  Anal Biochem        ISSN: 0003-2697            Impact factor:   3.365


  2 in total

1.  Characterization of a highly hop-resistant Lactobacillus brevis strain lacking hop transport.

Authors:  Jürgen Behr; Michael G Gänzle; Rudi F Vogel
Journal:  Appl Environ Microbiol       Date:  2006-10       Impact factor: 4.792

2.  Multicomponent quantification of Astragalus residue fermentation liquor using ion chromatography-integrated pulsed amperometric detection.

Authors:  Ying Zhang; Jiarong Wu; Quanhui Ni; Hong Dong
Journal:  Exp Ther Med       Date:  2017-06-26       Impact factor: 2.447

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.