Evaluation of the imaging properties of two generations of a CCD-based system for digital chest radiography.

Magnus Båth1, Patrik Sund, Lars Gunnar Månsson.   

Abstract

Two generations of a CCD-based detector system with lens-based optical coupling for digital chest radiography were evaluated in terms of presampling MTF, NPS, NEQ, DQE, linearity in response, and SNR over the detector area. Measurements were performed over a wide exposure range and at several different beam qualities. Neither the presampling MTF nor the DQE showed any general strong beam quality dependence, whereas the NPS and NEQ did when compared at specific entrance air kerma values. The exposure dependency for the DQE was found to be considerable, with the detectors showing low DQE at low exposures, and higher DQE at higher exposures. It was found that the second generation has been substantially improved compared to its predecessor regarding all the relevant parameters. The DQE(0) at an entrance air kerma of 5 microGy increased from 9% to 15%, mainly due to a better system gain (including optical coupling efficiency and matching of the energy of the emitted light photons to the sensitivity of the CCD camera). The first generation of detectors was found to have problems with bad peripheral resolution [MTF(muN/2) <0.1]. This problem was nonexistent for the second generation for which uniform resolution has been obtained [MTF(muN/2)=0.3]. A theoretical calculation of the DQE of two model systems similar to the ones evaluated was also performed, and the results were comparable to the experimentally determined data at high exposures. The model shows that both systems suffer from low optical coupling efficiency due to the large demagnification used. The main conclusion is that although the second generation has been improved, there is still a problem with low system gain leading to relatively modest DQE values, especially at low exposures.

Entities:  

Mesh:

Year:  2002        PMID: 12408303     DOI: 10.1118/1.1507781

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography.

Authors:  Patrik Sund; Magnus Båth; Susanne Kheddache; Lars Gunnar Månsson
Journal:  Eur Radiol       Date:  2003-10-16       Impact factor: 5.315

2.  Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practice.

Authors:  Wouter J H Veldkamp; Lucia J M Kroft; Mireille V Boot; Bart J A Mertens; Jacob Geleijns
Journal:  Eur Radiol       Date:  2005-08-31       Impact factor: 5.315

3.  Comparison of two detector systems for cone beam CT small animal imaging - a preliminary study.

Authors:  Yang Meng; Chris C Shaw; Xinming Liu; Mustafa C Altunbas; Tianpeng Wang; Lingyun Chen; Shu-Ju Tu; S Cheenu Kappadath; Chao-Jen Lai
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-03-02

4.  Correlation of the clinical and physical image quality in chest radiography for average adults with a computed radiography imaging system.

Authors:  C S Moore; T J Wood; A W Beavis; J R Saunderson
Journal:  Br J Radiol       Date:  2013-04-08       Impact factor: 3.039

Review 5.  Digital chest radiography: an update on modern technology, dose containment and control of image quality.

Authors:  Cornelia Schaefer-Prokop; Ulrich Neitzel; Henk W Venema; Martin Uffmann; Mathias Prokop
Journal:  Eur Radiol       Date:  2008-04-23       Impact factor: 5.315

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.