Literature DB >> 12406987

False-positive immunoassay results: a multicenter survey of erroneous immunoassay results from assays of 74 analytes in 10 donors from 66 laboratories in seven countries.

Vincent Marks1.   

Abstract

BACKGROUND: Analytical interference in immunoassays can produce serious errors, but it is generally considered rare with modern analytical systems.
METHOD: Blood was collected from 10 donors with illnesses known to be associated with rheumatoid factor. Immunoassays for 74 analytes were performed in 66 clinical laboratories. Each sample was measured in duplicate, and again in duplicate after the addition of a proprietary heterophil blocking reagent, with the laboratory's routinely used reagents and equipment. Reagents were typically supplied by the manufacturers of the closed analytical systems. Both competitive and sandwich immunoassays were investigated.
RESULTS: Overall approximately 8.7% of the 3445 results were considered to be "false positive", many of them seriously so. Twenty-one percent of the erroneously high results (1.8% of all results) were potentially misleading and were corrected by blocking reagent, although 49% of such seriously high results (4.2% of all results) were not. A further 39% of the false-positive results (2.6% of all results) would not necessarily have appeared likely to produce adverse clinical consequences but were substantially lowered by the addition of the blocking reagent.
CONCLUSIONS: Random errors can occur with all types of immunoassays tested and can be difficult to identify even when repeated in another laboratory. Clinicians need to be aware of these limitations.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12406987

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  14 in total

1.  Frequent false positive beta human chorionic gonadotropin tests in immunoglobulin A deficiency.

Authors:  A K Knight; T Bingemann; L Cole; C Cunningham-Rundles
Journal:  Clin Exp Immunol       Date:  2005-08       Impact factor: 4.330

2.  A common, under-recognized cause of elevated serum erythropoietin.

Authors:  Daria V Babushok; David Cardamone; Nichole Rulander; Stephen R Master; Elizabeth O Hexner
Journal:  Am J Med       Date:  2014-08-19       Impact factor: 4.965

3.  Steroid hormone levels in pregnancy and 1 year postpartum using isotope dilution tandem mass spectrometry.

Authors:  Offie P Soldin; Tiedong Guo; Elisabete Weiderpass; Rochelle E Tractenberg; Leena Hilakivi-Clarke; Steven J Soldin
Journal:  Fertil Steril       Date:  2005-09       Impact factor: 7.329

4.  Methods to identify saline-contaminated electrolyte profiles.

Authors:  Devin K Patel; Rishi D Naik; Richard B Boyer; John Wikswo; Eduard E Vasilevskis
Journal:  Clin Chem Lab Med       Date:  2015-09-01       Impact factor: 3.694

5.  False positive cardiac troponin elevation due to heterophile antibodies: more common than we recognise?

Authors:  Abbas Zaidi; Richard Cowell
Journal:  BMJ Case Rep       Date:  2010-07-15

6.  Interferences in immunoassay.

Authors:  Jill Tate; Greg Ward
Journal:  Clin Biochem Rev       Date:  2004-05

7.  Developing multiplexed assays for troponin I and interleukin-33 in plasma by peptide immunoaffinity enrichment and targeted mass spectrometry.

Authors:  Eric Kuhn; Terri Addona; Hasmik Keshishian; Michael Burgess; D R Mani; Richard T Lee; Marc S Sabatine; Robert E Gerszten; Steven A Carr
Journal:  Clin Chem       Date:  2009-04-16       Impact factor: 8.327

8.  Negative interference by rheumatoid factor of plasma B-type natriuretic peptide in chemiluminescent microparticle immunoassays.

Authors:  Wen Fan; Lei Xu; Liangcai Xie; Decai Yang; Xuezheng Liu; Jiajun Zhang; Yirong Li; Cunjian Yi
Journal:  PLoS One       Date:  2014-08-21       Impact factor: 3.240

9.  From radioimmunoassay to mass spectrometry: a new method to quantify orexin-A (hypocretin-1) in cerebrospinal fluid.

Authors:  Christophe Hirtz; Jérôme Vialaret; Audrey Gabelle; Nora Nowak; Yves Dauvilliers; Sylvain Lehmann
Journal:  Sci Rep       Date:  2016-05-11       Impact factor: 4.379

10.  Quality assessment of an interferon-gamma release assay for tuberculosis infection in a resource-limited setting.

Authors:  Nguyen T L Hang; Naoki Ishizuka; Naoto Keicho; Le T Hong; Do B Tam; Vu T X Thu; Ikumi Matsushita; Nobuyuki Harada; Kazue Higuchi; Shinsaku Sakurada; Luu T Lien
Journal:  BMC Infect Dis       Date:  2009-05-18       Impact factor: 3.090

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.