| Literature DB >> 12401363 |
Kazuyuki Ouchi1, Shaw Watanabe.
Abstract
Environmental exposure levels of bisphenol A (BPA) in human were investigated by measuring BPA glucuronide (BPA-G) in urine. After enzymatic hydrolysis of glucuronide substances in urine, BPA was extracted with diethyl ether. The extract was analyzed using a column-switching HPLC system employing a C(8) and a C(18) column with multi-channel coulometric electrochemical detection (ECD). The sensitivity and selectivity provided with redox mode of ECD allowed measurement of low level BPA in hydrolyzed urine. The quantification limit of BPA-G in urine was 0.2 ng/ml. RSDs of the intraassay precision were less than 3% and recoveries of the method were over 96% when analyzing BPA spiked urine samples (1.0 and 10 ng/ml). In a group of 48 women students, the level of BPA-G in urine ranged from 0.2 to 19.1 ng/ml with a median concentration of 1.2 ng/ml. Normalized against urinary creatinine, BPA-G ranged from 0.1 to 11.9 ng/mg creatinine with a median of 0.77 ng/mg creatinine. Copyright 2002 Elsevier Science B.V.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12401363 DOI: 10.1016/s1570-0232(02)00547-0
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205