What can we learn from explanted endovascular devices?

G Riepe1, C Heintz, E Kaiser, N Chakfé, M Morlock, M Delling, H Imig.   

Abstract

OBJECTIVE: To examine the durability of the Stentor and Vanguard endovascular devices in human implants.
METHODS: The textile covering, the polypropylene ligatures and the stent metal of 34 devices (25 Stentor, 9 Vanguard) with a mean duration of implantation of 28.8 +/- 16 months was examined by means of stereomicroscopy and scanning electron microscopy.
RESULTS: The polyester textile covering showed gaps along the sutured seam and isolated holes in the fabric. All of the examined polypropylene ligatures were worn, some ruptured. Four different types of stent corrosion were classified--pits (100%), bizarre craters (68%), large deficiencies (14%) and fractures (32%).
CONCLUSION: Holes in the polyester fabric and frame dislocations are specific for the design of Stentor and Vanguard grafts. The early corrosion of the stent metal Nitinol in these devices is surprising. Until more experience is gained with other devices, we have to be reminded, that the "gold standard" for the long-term durability of artificial vascular grafts is still "today's" conventional graft.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12389232     DOI: 10.1053/ejvs.2002.1677

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  5 in total

1.  Current practices in corrosion, surface characterization, and nickel leach testing of cardiovascular metallic implants.

Authors:  Srinidhi Nagaraja; Matthew Di Prima; David Saylor; Erica Takai
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-02-16       Impact factor: 3.368

2.  International guidelines for groin hernia management.

Authors: 
Journal:  Hernia       Date:  2018-01-12       Impact factor: 4.739

3.  Development of an optimized electrochemical process for subsequent coating of 316 stainless steel for stent applications.

Authors:  M Haïdopoulos; S Turgeon; C Sarra-Bournet; G Laroche; D Mantovani
Journal:  J Mater Sci Mater Med       Date:  2006-07       Impact factor: 4.727

4.  An in Vitro Twist Fatigue Test of Fabric Stent-Grafts Supported by Z-Stents vs. Ringed Stents.

Authors:  Jing Lin; Robert Guidoin; Jia Du; Lu Wang; Graeham Douglas; Danjie Zhu; Mark Nutley; Lygia Perron; Ze Zhang; Yvan Douville
Journal:  Materials (Basel)       Date:  2016-02-16       Impact factor: 3.623

5.  Intracavitary deposits on Essure® hysteroscopic sterilization devices: A case report.

Authors:  L W Maassen; D M van Gastel; E G W M Lentjes; M Y Bongers; S Veersema
Journal:  Case Rep Womens Health       Date:  2017-06-13
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.