Literature DB >> 12382737

Evaluating two extremity dosemeters based on LiF:Mg,Ti or LiF:Mg,Cu,P.

L Z Luo1, K J Velbeck, J E Rotunda.   

Abstract

Evaluation of a new extremity dosemeter is presented. The dosemeter is a passive device that is easy to wear and features a permanent individual numerical ID with barcode, a watertight case, an automatic TLD reader and database management software. Two dosemeters were studied: the first consists of a 100 mg x cm(-2) 7LiF:Mg,Ti (TLD-700) chip and a 42 mg x cm(-2) cap, the other consists of a 7 mg x cm(-2) layer of 7LiF:Mg,Cu,P (TLD-700H) powder and a 5 mg x cm(-2) cap. Sensitivity, repeatability, lower limit detection, angular responses and energy responses for these dosemeters are studied and presented. The dose calculation algorithm is developed and its dosimetric performance accuracy is compared with the standard ANSI N13.32-1995, Performance Testing of Extremity Dosemeters.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12382737     DOI: 10.1093/oxfordjournals.rpd.a005969

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Characterization of optically stimulated luminescence dosemeters to measure organ doses in diagnostic radiology.

Authors:  A Endo; T Katoh; I Kobayashi; R Joshi; J Sur; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2011-11-24       Impact factor: 2.419

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.