Thermodilution cardiac output. Cold vs room temperature injectate and the importance of measuring the injectate temperature in the right atrium.

P G Berthelsen1, N Eldrup, L B Nilsson, J P Rasmussen.   

Abstract

BACKGROUND: The feasibility of thermodilution cardiac output measurements with the more convenient room temperature thermal indicator instead of cold injectates has been repeatedly investigated. However, the issue has not been addressed with the appropriate statistical approach advocated by Altman and Bland. Furthermore, we wished to determine if the incorporation of a second thermistor in the thermodilution catheter, to measure the temperature of the thermal indicator where it is delivered into the right atrium/superior caval vein, would result in more precise cardiac output measurements.
METHODS: Fifty patients were randomized to receive a single or dual thermistor pulmonary artery thermodilution catheter. Cardiac output was calculated as the average of four injections of 10 ml of isotonic saline. Precision (2 x SD of differences in replicate measurements) for the two catheters and injectate temperatures, and bias and limits of agreement between measurements, with cold and room temperature injectates, were determined.
RESULTS: Precision was (0 degrees C) 0.42 l/min and (20 degrees C) 0.90 l/min, and bias and limits of agreement -0.83 l/min and -1.93-0.27 l/min for the single thermistor catheter. For the dual thermistor system precision was (0 degrees C) 0.34 l/min and (20 degrees C) 0.58 l/min. Bias and limits of agreement were -0.03 l/min and -0.61-0.55 l/min.
CONCLUSION: The second thermistor is redundant if cold injectates are used. If one wishes to use room temperature injectates the single thermistor system is inadequate. A dual thermistor catheter is, on the other hand, acceptable.

Entities:  

Mesh:

Year:  2002        PMID: 12366505     DOI: 10.1034/j.1399-6576.2002.460908.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  3 in total

1.  Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique.

Authors:  M Cecconi; D Dawson; R M Grounds; A Rhodes
Journal:  Intensive Care Med       Date:  2008-09-18       Impact factor: 17.440

Review 2.  Systematic review of cardiac output measurements by echocardiography vs. thermodilution: the techniques are not interchangeable.

Authors:  Mik Wetterslev; Hasse Møller-Sørensen; Rasmus Rothmann Johansen; Anders Perner
Journal:  Intensive Care Med       Date:  2016-03-01       Impact factor: 17.440

3.  Noninvasive Assessment of Cardiac Output: Accuracy and Precision of the Closed-Circuit Acetylene Rebreathing Technique for Cardiac Output Measurement.

Authors:  E Ashley Hardin; Douglas Stoller; Justin Lawley; Erin J Howden; Michinari Hieda; James Pawelczyk; Sara Jarvis; Kim Prisk; Satyam Sarma; Benjamin D Levine
Journal:  J Am Heart Assoc       Date:  2020-08-27       Impact factor: 5.501

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.