Literature DB >> 12352421

The mentor Alpha 1 penile prosthesis with reservoir lock-out valve: effective prevention of auto-inflation with improved capability for ectopic reservoir placement.

Steven K Wilson1, Gerard D Henry, John R Delk, Mario A Cleves.   

Abstract

PURPOSE: Auto-inflation is a common and annoying complication of 3-piece penile prostheses. In the published literature the rate is approximately 11% with a 2% operative revision rate. We report the results of a review of 160 Alpha 1 and NB implants (Mentor Corp., Santa Barbara, California) with the new lock-out valve located on the reservoir to treat impotence. We compared it with 339 Alpha 1 implants with the standard reservoir. We also investigated the lock-out reservoir for ectopic nonretropubic implantation.
MATERIALS AND METHODS: We compared 339 Alpha prostheses with the standard reservoir that were implanted between January 1, 1998 through December 31, 1999 and 160 with the new lock-out valve placed since January 2000 with at least 6 months of followup. Implants were further stratified as first time (virgin) or revision-replacement of a previous implant. In 8 patients with a scarred or obliterated retroperitoneal space the lock-out reservoir was placed superior to the transversalis fascia and beneath the abdominal musculature.
RESULTS: Kaplan-Meier estimated 1-year survival was not significantly different in terms of mechanical failure (p = 0.57 and 0.85) revision for any cause (p = 0.92 and 0.92), patient dissatisfaction (p = 0.35 and 0.11) or infection (p = 0.64 and 0.94) for all implants and virgin implants only, respectively. Only 2 patients (1.3%) with a lock-out valve complained of auto-inflation initially and the problem resolved in each after instruction on how to operate the device. Of the patients in the earlier series 11% complained of auto-inflation and 2% required operative correction. None of the 8 patients with an ectopic reservoir location complained of auto-inflation.
CONCLUSIONS: Our results indicate that the lock-out valve prevents early auto-inflation. Addition of the lock-out valve does not impact the revision rate compared with the same implant with a standard reservoir. In patients with a scarred retropubic space the lock-out valve offers the penile implant surgeon a decreased probability of auto-inflation with ectopic reservoir placement.

Entities:  

Mesh:

Year:  2002        PMID: 12352421     DOI: 10.1097/01.ju.0000030603.29170.c2

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  23 in total

Review 1.  A practical overview of considerations for penile prosthesis placement.

Authors:  Landon Trost; Philip Wanzek; George Bailey
Journal:  Nat Rev Urol       Date:  2015-12-01       Impact factor: 14.432

Review 2.  Penile prosthetic surgery and its role in the treatment of end-stage erectile dysfunction - an update.

Authors:  S Jain; T R Terry
Journal:  Ann R Coll Surg Engl       Date:  2006-07       Impact factor: 1.891

3.  Incarcerated inguinal hernia and small bowel obstruction as a rare complication of a penile prosthesis.

Authors:  S J Serio; P Schafer; A M Merchant
Journal:  Hernia       Date:  2012-09-26       Impact factor: 4.739

Review 4.  Alternative Reservoir and Balloon Placement in Complex Patients Undergoing Urologic Prostheses.

Authors:  Travis J Pagliara; Daniel W Smith; Boyd R Viers; Allen F Morey
Journal:  Curr Urol Rep       Date:  2018-04-13       Impact factor: 3.092

Review 5.  Innovating Incrementally: Development of the Modern Inflatable Penile Prosthesis.

Authors:  Mark Ehlers; Benjamin McCormick; R Matthew Coward; Bradley D Figler
Journal:  Curr Urol Rep       Date:  2019-03-07       Impact factor: 3.092

6.  Longitudinal and Horizontal Load Testing of Inflatable Penile Implant Cylinders of Two Manufacturers: An Ex Vivo Demonstration of Inflated Rigidity.

Authors:  Jason M Scovell; Liehui Ge; Enrique V Barrera; Steven K Wilson; Rafael E Carrion; Tariq S Hakky
Journal:  J Sex Med       Date:  2016-11       Impact factor: 3.802

Review 7.  Reservoir Placement Considerations During Inflatable Penile Prosthesis Surgery.

Authors:  Steven J Hudak
Journal:  Curr Urol Rep       Date:  2019-01-28       Impact factor: 3.092

Review 8.  The evolution of the inflatable penile prosthetic device and surgical innovations with anatomical considerations.

Authors:  Tariq S Hakky; Run Wang; Gerard D Henry
Journal:  Curr Urol Rep       Date:  2014-06       Impact factor: 3.092

Review 9.  A comparative review of the options for treatment of erectile dysfunction: which treatment for which patient?

Authors:  Konstantinos Hatzimouratidis; Dimitrios G Hatzichristou
Journal:  Drugs       Date:  2005       Impact factor: 9.546

Review 10.  Current management of penile implant infections, device reliability, and optimizing cosmetic outcome.

Authors:  John J Mulcahy; Andrew Kramer; William O Brant; Justin L Parker; Paul E Perito; Jeremy B Myers; Richard Bryson; Meagan Dunne
Journal:  Curr Urol Rep       Date:  2014-06       Impact factor: 3.092

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.