Literature DB >> 12350475

Laparoscopic linear cutting stapler failure.

Donna Y Deng1, Maxwell V Meng, Hiep T Nguyen, Gary C Bellman, Marshall L Stoller.   

Abstract

OBJECTIVES: To characterize the frequency and nature of problems with linear cutting staplers to help prevent complications in the future. These devices are often used during laparoscopic urologic procedures.
METHODS: We retrospectively reviewed the experience with laparoscopic linear cutting staplers at two institutions routinely performing urologic laparoscopy and analyzed the difficulties with any staplers. Data from the Food and Drug Administration Center for Devices and Radiological Health were also examined to determine the prevalence and types of reported problems.
RESULTS: In performing approximately 460 laparoscopic cases, we encountered 5 problems (1%) with endovascular gastrointestinal anastomosis staplers. Fifty-five additional cases in 50 patients were documented in the Food and Drug Administration database. Of the 55 patients, 15 (27%) required open conversion to manage the problem, 8 (15%) received blood transfusions, and 2 (4%) died postoperatively. Twenty-two events occurred during 19 laparoscopic donor nephrectomies (35%) without associated graft dysfunction, damage, or loss. All phases of instrument use were subject to problems; however, abnormal firing of the stapler and improper staple formation were the most common and morbid aspects of device malfunction.
CONCLUSIONS: Despite the general reliability of linear cutting staplers, difficulties were encountered in every step of use. Most situations were successfully managed by prompt identification and appropriate intracorporeal maneuvers. Nevertheless, significant morbidity may occur, and conversion to an open operation should be considered. Many potential problems can be avoided by surgeon and staff education, and one should be aware of the alternative methods of tissue ligation currently available.

Entities:  

Mesh:

Year:  2002        PMID: 12350475     DOI: 10.1016/s0090-4295(02)01778-8

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  12 in total

Review 1.  [Ten years of laparoscopic living kidney donation. From an extravagant to a routine procedure].

Authors:  M Giessing; T F Fuller; S Deger; J Roigas; M Tüllmann; L Liefeldt; K Budde; T Fischer; B Winkelmann; D Schnorr; S A Loening
Journal:  Urologe A       Date:  2006-01       Impact factor: 0.639

2.  Fatal and nonfatal hemorrhagic complications of living kidney donation.

Authors:  Hervé Baumert
Journal:  Ann Surg       Date:  2006-10       Impact factor: 12.969

3.  Dangers in the use of staplers in liver surgery.

Authors:  Angela Riga; Nariman Karanjia
Journal:  Ann R Coll Surg Engl       Date:  2007-11       Impact factor: 1.891

4.  Application of the Hem-o-lok ligation system in laparoscopic nephrectomy.

Authors:  Hao Ping; Nian-Zeng Xing; Jun-Hui Zhang; Yong Yan; Ning Kang; Yi-Nong Niu
Journal:  Surg Endosc       Date:  2009-12-24       Impact factor: 4.584

5.  Laparoscopic donor nephrectomy, complications and management: a single center experience.

Authors:  Volkan Tuğcu; Selçuk Şahin; İsmail Yiğitbaşı; Nevzat Can Şener; Fatih Gökhan Akbay; Ali İhsan Taşçı
Journal:  Turk J Urol       Date:  2017-01-06

Review 6.  Evolution of surgical skills training.

Authors:  Kurt-E Roberts; Robert-L Bell; Andrew-J Duffy
Journal:  World J Gastroenterol       Date:  2006-05-28       Impact factor: 5.742

Review 7.  Staplers or clips?: A systematic review and meta-analysis of vessel controlling devices for renal pedicle ligation in laparoscopic live donor nephrectomy.

Authors:  Yu Liu; Zhongli Huang; Yuntian Chen; Banghua Liao; Deyi Luo; Xiaoshuai Gao; Kunjie Wang; Hong Li
Journal:  Medicine (Baltimore)       Date:  2018-11       Impact factor: 1.889

8.  A comparison of renal vascular control techniques during laparoscopic nephrectomy.

Authors:  Gokhan Koc; Gokhan Rahmi Ekin; Batuhan Ergani; Yusuf Ozlem Ilbey
Journal:  J Minim Access Surg       Date:  2021 Apr-Jun       Impact factor: 1.407

Review 9.  Surgical complications of laparoscopic urological surgery.

Authors:  Michael S Lasser; Reza Ghavamian
Journal:  Arab J Urol       Date:  2012-01-09

10.  Endo GIA stapler malfunction in a small bowel loop resection.

Authors:  Vinu Perinjelil; Kwabena Nkansah-Amankra; Tareq Maraqa; Leo Mercer; Gul Sachwani-Daswani
Journal:  J Surg Case Rep       Date:  2018-10-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.