Christian Neusüss1, Matthias Pelzing, Marcus Macht.
Abstract
A capillary electrophoresis-tandem mass spectrometry (CE-MS/MS) approach has been developed for routine application in proteomic studies. Robustness of the coupling is achieved by using a standard coaxial sheath-flow sprayer. Thereby, greater stability than nanoelectrospray ionization-mass spectrometry coupling of sheathless capillary electrophoresis or nanoliquid chromatography (nano-LC) is achieved, resulting in stable operation for several weeks and unattended overnight sequences. The applied sheath flow is reduced to 1-2 microL/min in order to increase sensitivity. Standard peptides and those of digests of standard proteins and gel-separated proteins can be detected in the low femtomole range (full scan and MS/MS). Detection limits are found to be as low as 500 amol. Low femtomole amounts are required for unequivocal identification by MS/MS experiments in the ion trap and subsequent database search. By applying a simple pH-mediated stacking the concentration sensitivity can be lowered to some tens of fmol/microL (nM), depending on capillary size. This sensitivity is close to published values for sheathless CE-MS and nano-LC-MS, respectively (a comparison to reference values is presented). Moreover, with capillaries of about 50 cm in length separations in less than 10 min are possible resulting in a throughput of up to four analyses per hour. This is a factor of 4-12 times faster than nano-LC separation, being the state-of-the-art techniques for proteomic studies.Mesh:
Substances:
Year: 2002 PMID: 12298087 DOI: 10.1002/1522-2683(200209)23:18<3149::AID-ELPS3149>3.0.CO;2-8
Source DB: PubMed Journal: Electrophoresis ISSN: 0173-0835 Impact factor: 3.535