| Literature DB >> 12296336 |
Kurt P Tschopp1, Christine Gottardo.
Abstract
In the present study, 3 types of electrodes for recurrent laryngeal nerve (RLN) monitoring are compared: 1) intralaryngeal surface electrodes attached to a conventional endotracheal tube, 2) monopolar needle electrodes placed on the vocal cords by direct laryngoscopy, and 3) bipolar needle electrodes inserted intraoperatively through the cricothyroid ligament. Data were collected from stimulation of 21 RLNs in 16 patients undergoing thyroid surgery. The reliability in terms of distinct electromyographic (EMG) potentials following stimulation of the RLN was 100% with monopolar and bipolar needle electrodes and 76% with intralaryngeal surface electrodes. The mean (+/-SD) amplitudes of the EMG potentials were 1.61 +/- 1.6 mV, 2.37 +/- 1.8 mV, and 0.35 +/- 0.4 mV for the monopolar endolaryngeal, bipolar transligament, and intralaryngeal surface electrodes, respectively. The advantages and disadvantages of each type of electrode are discussed.Entities:
Mesh:
Year: 2002 PMID: 12296336 DOI: 10.1177/000348940211100909
Source DB: PubMed Journal: Ann Otol Rhinol Laryngol ISSN: 0003-4894 Impact factor: 1.547