Literature DB >> 18305996

Intraoperative monitoring of the recurrent laryngeal nerve in thyroid surgery.

H Dralle1, C Sekulla, K Lorenz, M Brauckhoff, A Machens.   

Abstract

BACKGROUND: Recurrent laryngeal nerve (RLN) palsy ranks among the leading reasons for medicolegal litigation of surgeons because of its attendant reduction in quality of life. As a risk minimization tool, intraoperative nerve monitoring (IONM) has been introduced to verify RLN function integrity intraoperatively. Nevertheless, a systematic evidence-based assessment of this novel health technology has not been performed.
METHODS: The present study was based on a systematic appraisal of the literature using evidence-based criteria.
RESULTS: Recurrent laryngeal nerve palsy rates (RLNPR) varied widely after thyroid surgery, ranging from 0%-7.1% for transient RLN palsy to 0%-11% for permanent RLN palsy. These rates did not differ much from those reported for visual nerve identification without the use of IONM. Six studies with more than 100 nerves at risk (NAR) each evaluated RLNPR by contrasting IONM with visual nerve identification only. Recurrent laryngeal nerve palsy rates tended to be lower with IONM than without it, but this difference was not statistically significant. Six additional studies compared IONM findings with their corresponding postoperative laryngoscopic results. Those studies revealed high negative predictive values (NPV; 92%-100%), but relatively low and variable positive predictive values (PPV; 10%-90%) for IONM, limiting its utility for intraoperative RLN management.
CONCLUSIONS: Apart from navigating the surgeon through challenging anatomies, IONM may lend itself as a routine adjunct to the gold standard of visual nerve identification. To further reduce the number of false negative IONM signals, the causes underlying its relatively low PPV require additional clarification.

Entities:  

Mesh:

Year:  2008        PMID: 18305996     DOI: 10.1007/s00268-008-9483-2

Source DB:  PubMed          Journal:  World J Surg        ISSN: 0364-2313            Impact factor:   3.352


  74 in total

1.  Intraoperative electromyogram monitoring of the recurrent laryngeal nerve: experience with an intralaryngeal surface electrode. A method to reduce the risk of recurrent laryngeal nerve injury during thyroid surgery.

Authors:  D Horn; V M Rötzscher
Journal:  Langenbecks Arch Surg       Date:  1999-08       Impact factor: 3.445

2.  Identification of the non-recurrent inferior laryngeal nerve using intraoperative neurostimulation.

Authors:  Michael Brauckhoff; Gerard Walls; Katrin Brauckhoff; Phuong Nguyen Thanh; Oliver Thomusch; Henning Dralle
Journal:  Langenbecks Arch Surg       Date:  2001-10-27       Impact factor: 3.445

3.  Recurrent laryngeal nerve identification and assessment during thyroid surgery: laryngeal palpation.

Authors:  Gregory W Randolph; James B Kobler; Jamie Wilkins
Journal:  World J Surg       Date:  2004-08-03       Impact factor: 3.352

4.  Comparison of various methods of electromyographic monitoring of the recurrent laryngeal nerve in thyroid surgery.

Authors:  Kurt P Tschopp; Christine Gottardo
Journal:  Ann Otol Rhinol Laryngol       Date:  2002-09       Impact factor: 1.547

5.  Intraoperative recurrent laryngeal nerve monitoring.

Authors:  D H Rice; B Cone-Wesson
Journal:  Otolaryngol Head Neck Surg       Date:  1991-09       Impact factor: 3.497

6.  Intraoperative identification of the recurrent laryngeal nerve and demonstration of its function.

Authors:  D J Premachandra; G J Radcliffe; M P Stearns
Journal:  Laryngoscope       Date:  1990-01       Impact factor: 3.325

7.  Continuous laryngeal nerve integrity monitoring during thyroidectomy: does it reduce risk of injury?

Authors:  Matt L Robertson; David L Steward; Jack L Gluckman; Jeff Welge
Journal:  Otolaryngol Head Neck Surg       Date:  2004-11       Impact factor: 3.497

8.  12-h pretreatment with methylprednisolone versus placebo for prevention of postextubation laryngeal oedema: a randomised double-blind trial.

Authors:  Bruno François; Eric Bellissant; Valérie Gissot; Arnaud Desachy; Sandrine Normand; Thierry Boulain; Olivier Brenet; Pierre-Marie Preux; Philippe Vignon
Journal:  Lancet       Date:  2007-03-31       Impact factor: 79.321

9.  Laryngeal nerve identification during thyroid surgery -- feasibility of a novel approach.

Authors:  Carl L Hillermann; Joe Tarpey; David E Phillips
Journal:  Can J Anaesth       Date:  2003-02       Impact factor: 5.063

10.  Incidence of vocal cord paralysis with and without recurrent laryngeal nerve monitoring during thyroidectomy.

Authors:  Maisie Shindo; Neil N Chheda
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2007-05
View more
  120 in total

1.  Visualization versus neuromonitoring of recurrent laryngeal nerves during thyroidectomy: what about the costs?

Authors:  Gianlorenzo Dionigi; Alessandro Bacuzzi; Luigi Boni; Stefano Rausei; Francesca Rovera; Renzo Dionigi
Journal:  World J Surg       Date:  2012-04       Impact factor: 3.352

Review 2.  [Intraoperative complications of neck surgery].

Authors:  D Simon; M Lassau; P Schmidt-Wilcke; M Boucher
Journal:  Chirurg       Date:  2012-07       Impact factor: 0.955

3.  [Current thyroid and parathyroid surgery].

Authors:  H Dralle
Journal:  Chirurg       Date:  2010-07       Impact factor: 0.955

4.  [Intraoperative neuromonitoring of thyroid gland operations : Surgical standards and aspects of expert assessment].

Authors:  H Dralle; K Lorenz
Journal:  Chirurg       Date:  2010-07       Impact factor: 0.955

Review 5.  Clinical guidelines on intraoperative neuromonitoring during thyroid and parathyroid surgery.

Authors:  Hui Sun; Wen Tian; Kewei Jiang; Fengyu Chiang; Ping Wang; Tao Huang; Jingqiang Zhu; Jianwu Qin; Xiaoli Liu
Journal:  Ann Transl Med       Date:  2015-09

6.  Severity of Recurrent Laryngeal Nerve Injuries in Thyroid Surgery.

Authors:  Gianlorenzo Dionigi; Che-Wei Wu; Hoon Yub Kim; Stefano Rausei; Luigi Boni; Feng-Yu Chiang
Journal:  World J Surg       Date:  2016-06       Impact factor: 3.352

7.  Recurrent laryngeal nerve injury with incomplete loss of electromyography signal during monitored thyroidectomy-evaluation and outcome.

Authors:  Che-Wei Wu; Min Hao; Mengzi Tian; Gianlorenzo Dionigi; Ralph P Tufano; Hoon Yub Kim; Kwang Yoon Jung; Xiaoli Liu; Hui Sun; I-Cheng Lu; Pi-Ying Chang; Feng-Yu Chiang
Journal:  Langenbecks Arch Surg       Date:  2016-02-17       Impact factor: 3.445

8.  Five-year follow-up of a randomized clinical trial of total thyroidectomy versus Dunhill operation versus bilateral subtotal thyroidectomy for multinodular nontoxic goiter.

Authors:  Marcin Barczyński; Aleksander Konturek; Alicja Hubalewska-Dydejczyk; Filip Gołkowski; Stanisław Cichoń; Wojciech Nowak
Journal:  World J Surg       Date:  2010-06       Impact factor: 3.352

9.  Evidence-based review series on endoscopic thyroidectomy: real progress and future trends.

Authors:  Gianlorenzo Dionigi
Journal:  World J Surg       Date:  2009-02       Impact factor: 3.352

10.  Commentary on transoral access for endoscopic thyroid resection : Witzel K, von Rahden BH, Kaminski C, Stein HJ (2008) Transoral access for endoscopic thyroid resection. Surg Endosc 22(8):1871-1875.

Authors:  Gianlorenzo Dionigi; Francesca Rovera; Luigi Boni
Journal:  Surg Endosc       Date:  2008-12-06       Impact factor: 4.584

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.