Literature DB >> 12216729

Plasticizer contamination from vacuum system O-rings in a quadrupole ion trap mass spectrometer.

Kent M Verge1, George R Agnes.   

Abstract

The outgassing of plasticizers from Buna-N and Viton o-rings under vacuum lead to undesired ion-molecule chemistry in an Electrospray Quadrupole Ion Trap Mass Spectrometer. In experiments with the helium bath gas pressure >1.2 mTorr, or whenever analyte ions were stored for >100 ms, extensive loss of analyte ions by proton transfer or adduction with o-ring plasticizers bis(2-ethylhexyl) phthalate and bis(2-ethylhexyl) adipate occurred. A temporary solution to this contamination problem was found to be overnight refluxing in hexane of all the o-rings in the vacuum system. This procedure alleviated this plasticizer contamination for approximately 100 hours of operation. These results, and those that lead to identification of the contamination as plasticizers outgassing from o-rings are described.

Entities:  

Year:  2002        PMID: 12216729     DOI: 10.1016/S1044-0305(02)00386-0

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  7 in total

1.  Chemicals in laboratory room air stimulate olfactory neurons of female Bombyx mori.

Authors:  J Ziesmann; I Valterova; K Haberkorn; M G de Brito Sanchez; K Kaissling
Journal:  Chem Senses       Date:  2000-02       Impact factor: 3.160

2.  A combined linear ion trap time-of-flight system with improved performance and MS(n) capabilities.

Authors:  B A Collings; J M Campbell; D Mao; D J Douglas
Journal:  Rapid Commun Mass Spectrom       Date:  2001       Impact factor: 2.419

3.  High resolution on a quadrupole ion trap mass spectrometer.

Authors:  J C Schwartz; J E Syka; I Jardine
Journal:  J Am Soc Mass Spectrom       Date:  1991-05       Impact factor: 3.109

4.  Plasticizers in pacifiers: direct determination by FAB-MS.

Authors:  J O Lay; B J Miller
Journal:  Anal Chem       Date:  1987-11-15       Impact factor: 6.986

Review 5.  Ion trap mass spectrometry. Using high-pressure ionization.

Authors:  S A McLuckey; G J Van Berkel; D E Goeringer; G L Glish
Journal:  Anal Chem       Date:  1994-07-15       Impact factor: 6.986

6.  Estimation of the environmental contamination by phthalic acid esters leaching from household wastes.

Authors:  M J Bauer; R Herrmann
Journal:  Sci Total Environ       Date:  1997-12-03       Impact factor: 7.963

7.  Poly(ethylene glycol) doubly and singly cationized by different alkali metal ions: relative cation affinities and cation-dependent resolution in a quadrupole ion trap mass spectrometer.

Authors:  Michael J Bogan; George R Agnes
Journal:  J Am Soc Mass Spectrom       Date:  2002-02       Impact factor: 3.109

  7 in total
  9 in total

1.  Identification of a new source of interference leached from polypropylene tubes in mass-selective analysis.

Authors:  Yuan-Qing Xia; Shefali Patel; Ray Bakhtiar; Ronald B Franklin; George A Doss
Journal:  J Am Soc Mass Spectrom       Date:  2005-03       Impact factor: 3.109

2.  Rapid Assessment of Contaminants and Interferences in Mass Spectrometry Data Using Skyline.

Authors:  Matthew J Rardin
Journal:  J Am Soc Mass Spectrom       Date:  2018-04-17       Impact factor: 3.109

3.  Formation of iron complexes from trifluoroacetic acid based liquid chromatography mobile phases as interference ions in liquid chromatography/electrospray ionization mass spectrometric analysis.

Authors:  Anil Shukla; Rui Zhang; Daniel J Orton; Rui Zhao; Therese R W Clauss; Ronald Moore; Richard D Smith
Journal:  Rapid Commun Mass Spectrom       Date:  2011-05-30       Impact factor: 2.419

4.  Anion-Anion Chemistry with Mass-Selected Molecular Fragments on Surfaces.

Authors:  Fangshun Yang; K Antonio Behrend; Harald Knorke; Markus Rohdenburg; Ales Charvat; Carsten Jenne; Bernd Abel; Jonas Warneke
Journal:  Angew Chem Int Ed Engl       Date:  2021-10-13       Impact factor: 16.823

5.  Identification of diethylene glycol monobutyl ether as a source of contamination in an ion trap mass spectrometer.

Authors:  Christopher R Gibson; Cheri M Brown
Journal:  J Am Soc Mass Spectrom       Date:  2003-11       Impact factor: 3.109

6.  Cyclic polyamide oligomers extracted from nylon 66 membrane filter disks as a source of contamination in liquid chromatography/mass spectrometry.

Authors:  John C Tran; Alan A Doucette
Journal:  J Am Soc Mass Spectrom       Date:  2006-03-06       Impact factor: 3.109

7.  Liquid chromatography-tandem mass spectrometry determination of plasma 24S-hydroxycholesterol with chromatographic separation of 25-hydroxycholesterol.

Authors:  Andrea E DeBarber; Dieter Lütjohann; Louise Merkens; Robert D Steiner
Journal:  Anal Biochem       Date:  2008-06-13       Impact factor: 3.365

8.  Identification of dimethyldioctadecylammonium ion (m/z 550.6) and related species (m/z 522.6, 494.6) as a source of contamination in mass spectrometry.

Authors:  M Lisa Manier; D Shannon Cornett; David L Hachey; Richard M Caprioli
Journal:  J Am Soc Mass Spectrom       Date:  2008-02-08       Impact factor: 3.109

9.  Self-organizing layers from complex molecular anions.

Authors:  Jonas Warneke; Martin E McBriarty; Shawn L Riechers; Swarup China; Mark H Engelhard; Edoardo Aprà; Robert P Young; Nancy M Washton; Carsten Jenne; Grant E Johnson; Julia Laskin
Journal:  Nat Commun       Date:  2018-05-14       Impact factor: 14.919

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.