| Literature DB >> 12216729 |
Kent M Verge1, George R Agnes.
Abstract
The outgassing of plasticizers from Buna-N and Viton o-rings under vacuum lead to undesired ion-molecule chemistry in an Electrospray Quadrupole Ion Trap Mass Spectrometer. In experiments with the helium bath gas pressure >1.2 mTorr, or whenever analyte ions were stored for >100 ms, extensive loss of analyte ions by proton transfer or adduction with o-ring plasticizers bis(2-ethylhexyl) phthalate and bis(2-ethylhexyl) adipate occurred. A temporary solution to this contamination problem was found to be overnight refluxing in hexane of all the o-rings in the vacuum system. This procedure alleviated this plasticizer contamination for approximately 100 hours of operation. These results, and those that lead to identification of the contamination as plasticizers outgassing from o-rings are described.Entities:
Year: 2002 PMID: 12216729 DOI: 10.1016/S1044-0305(02)00386-0
Source DB: PubMed Journal: J Am Soc Mass Spectrom ISSN: 1044-0305 Impact factor: 3.109