Validation of the A&D UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance.

Daniele Longo1, Olivo Bertolo, Gianluca Toffanin, Paola Frezza, Paolo Palatini.   

Abstract

OBJECTIVES: To determine the accuracy of the UA-631 (UA-779 Life Source for the American market) blood pressure monitor developed by the A&D Company (Toshimi-ku, Tokyo, Japan).
DESIGN: Device evaluation was performed using a new protocol proposed by the Working Group on blood pressure monitoring of the European Society of Hypertension (ESH). Monitor performance was assessed in relation to subjects' gender, age, skinfold thickness, arm circumference, BMI, and elasticity index of large (C1) and small (C2) arteries.
METHODS: The A&D recorder was assessed according to the various phases of the protocol. Sequential readings were taken for the main validation test. Outcome was classified according to the criteria of ESH recommendations, which are based on four zones of accuracy differing from the mercury standard by 5, 10, 15 mmHg, or more.
RESULTS: The main validation test was performed in 66 subjects for a total of 198 device measurements. The A&D monitor passed all three phases both for systolic and diastolic blood pressure (SBP and DBP). Mean blood pressure difference between device and observers was 2 +/- 5 mmHg for SBP and 1 +/- 3 mmHg for DBP. The absolute discrepancy between device and observers (4 +/- 4 mmHg for SBP, and 2 +/- 2 mmHg for DBP) was related to age (negatively) and to C1 (positively), but in a multivariable regression analysis only C1 remained a significant independent predictor of the absolute device-observer discrepancy.
CONCLUSIONS: These data show that the A&D UA-631 device satisfies the new recommended ESH accuracy levels for both SBP and DBP. Its performance seems to be better in subjects with stiffer arteries. Copyright 2002 Lippincott Williams & Wilkins

Entities:  

Mesh:

Year:  2002        PMID: 12198341     DOI: 10.1097/00126097-200208000-00007

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  6 in total

1.  Difference in blood pressure readings with mercury and automated devices: Impact on hypertension prevalence estimates in Dar es Salaam, Tanzania.

Authors:  Arnaud Chiolero; Jean-Pierre Gervasoni; Anne Rwebogora; Marianna Balampama; Fred Paccaud; Pascal Bovet
Journal:  Eur J Epidemiol       Date:  2006-07-07       Impact factor: 8.082

2.  Effects of sex and hypertension subtype on haemodynamics and left ventricular diastolic function in older patients with stage 1 hypertension.

Authors:  Naoki Fujimoto; Yoshiyuki Okada; Shigeki Shibata; Stuart A Best; Tiffany B Bivens; Benjamin D Levine; Qi Fu
Journal:  J Hypertens       Date:  2013-11       Impact factor: 4.844

3.  Masked hypertension defined by ambulatory blood pressure monitoring is associated with an increased serum glucose level and urinary albumin-creatinine ratio.

Authors:  Joji Ishikawa; Satoshi Hoshide; Kazuo Eguchi; Joseph E Schwartz; Thomas G Pickering; Kazuyuki Shimada; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2010-08       Impact factor: 3.738

4.  Creatine kinase as predictor of blood pressure and hypertension. Is it all about body mass index? A follow-up study of 250 patients.

Authors:  Stein H Johnsen; Hallvard Lilleng; Svein I Bekkelund
Journal:  J Clin Hypertens (Greenwich)       Date:  2014-10-03       Impact factor: 3.738

5.  Association between Anthropometric Measures and Indicators for Hypertension Control among Kazakh-Chinese Hypertension Patients in Xinjiang, China: Results from a Cross-sectional Study.

Authors:  Qinghua Zhang; Tanmay Mahapatra; Feifei Huang; Weiming Tang; Yufang Guo; Songyuan Tang; Yang Lei; Lei Feng; Anni Wang; Liuyi Zhang; Jingping Zhang
Journal:  PLoS One       Date:  2017-01-27       Impact factor: 3.240

6.  Association between blood pressure categories and cardiovascular disease mortality in China.

Authors:  Jie Guo; Jun Lv; Yu Guo; Zheng Bian; Bang Zheng; Man Wu; Ling Yang; Yiping Chen; Jian Su; Jianqiang Zhang; Jvying Yao; Junshi Chen; Zhengming Chen; Canqing Yu; Liming Li
Journal:  PLoS One       Date:  2021-07-30       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.