Literature DB >> 12169969

Variables affecting test-retest variability of Heidelberg Retina Tomograph II stereometric parameters.

Ramanjit Sihota1, Vikas Gulati, Harish C Agarwal, Rohit Saxena, Ajay Sharma, Ravindra M Pandey.   

Abstract

PURPOSE: To study the test-retest variability of stereometric parameters on the Heidelberg Retina Tomograph II, a new clinical instrument for glaucoma management.
METHODS: In a cross-sectional study of 24 consecutive cases of glaucoma and 26 healthy subjects, Heidelberg Retina Tomograph II stereometric parameters from five consecutive images were obtained for one randomly selected eye of each subject. Test-retest variability was studied using three different statistical methods (coefficient of variance, intraclass correlation coefficient, and Cronbach alpha). The effect of age, diagnosis, linear cup/disc ratio, visual acuity, and refractive error on test-retest variability on HRT II was analyzed. The repeatability of Moorfields regression analysis and the baseline variability in the progression analysis software was also studied.
RESULTS: Using coefficient of variance, intraclass correlation coefficient, and Cronbach alpha, the test-retest variability was found to be lowest for mean cup depth, cup area, cup/disc area ratio, vertical cup/disc ratio and rim/disc area ratio, in that order. Test-retest variability had a significant correlation with age (r = 0.33, P = 0.019) and visual acuity (r = -0.46, P = 0.005). Compared with eyes with astigmatism less than 1 D (mean coefficient of variance = 6.4 +/- 4.9), the test-retest variability was higher (P = 0.044) in eyes with astigmatism more than 1D (mean coefficient of variance = 20.0 +/- 22.6). Moorfields regression analysis was inconsistent in 52% cases. The average baseline change in progression analysis software was 0.076 +/- 0.081.
CONCLUSION: The test-retest variability of Heidelberg Retina Tomograph II stereometric parameters is comparable to that reported for the Heidelberg Retina Tomograph. Eyes with uncorrected astigmatism more than 1 D and poor visual acuity may have a higher variability of Heidelberg Retina Tomograph II stereometric parameters.

Entities:  

Mesh:

Year:  2002        PMID: 12169969     DOI: 10.1097/00061198-200208000-00008

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  15 in total

1.  Comparison of long-term fluctuations: laser scanning tomography versus automated perimetry.

Authors:  Jens Funk; Heiko Mueller
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-08-29       Impact factor: 3.117

Review 2.  Imaging in glaucoma.

Authors:  Daniel M Stein; Gadi Wollstein; Joel S Schuman
Journal:  Ophthalmol Clin North Am       Date:  2004-03

3.  Improving the repeatability of Heidelberg retina tomograph and Heidelberg retina tomograph II rim area measurements.

Authors:  N G Strouthidis; E T White; V M F Owen; T A Ho; D F Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

4.  Detecting glaucoma with RADAAR: the Bridlington Eye Assessment Project.

Authors:  M J Hawker; S A Vernon; C L Tattersall; H S Dua
Journal:  Br J Ophthalmol       Date:  2006-06       Impact factor: 4.638

Review 5.  Test-retest variability in structural parameters measured with glaucoma imaging devices.

Authors:  Makoto Araie
Journal:  Jpn J Ophthalmol       Date:  2012-11-09       Impact factor: 2.447

6.  Agreement in assessing cup-to-disc ratio measurement among stereoscopic optic nerve head photographs, HRT II, and Stratus OCT.

Authors:  Stella N Arthur; Aric J Aldridge; Julio De León-Ortega; Gerald McGwin; Aiyuan Xie; Christopher A Girkin
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

7.  Advanced imaging for glaucoma study: design, baseline characteristics, and inter-site comparison.

Authors:  Phuc V Le; Xinbo Zhang; Brian A Francis; Rohit Varma; David S Greenfield; Joel S Schuman; Nils Loewen; David Huang
Journal:  Am J Ophthalmol       Date:  2014-11-08       Impact factor: 5.258

Review 8.  'Structure-function relationship' in glaucoma: past thinking and current concepts.

Authors:  Rizwan Malik; William H Swanson; David F Garway-Heath
Journal:  Clin Exp Ophthalmol       Date:  2012-04-12       Impact factor: 4.207

9.  Effect of glaucomatous damage on repeatability of confocal scanning laser ophthalmoscope, scanning laser polarimetry, and optical coherence tomography.

Authors:  Julio E DeLeón Ortega; Lisandro M Sakata; Bobby Kakati; Gerald McGwin; Blythe E Monheit; Stella N Arthur; Christopher A Girkin
Journal:  Invest Ophthalmol Vis Sci       Date:  2007-03       Impact factor: 4.799

10.  New nonlinear multivariable model shows the relationship between central corneal thickness and HRTII topographic parameters in glaucoma patients.

Authors:  Dimitrios Kourkoutas; Gerasimos Georgopoulos; Antonios Maragos; Ioannis Apostolakis; George Tsekouras; Irene S Karanasiou; Dimitrios Papaconstantinou; Evaggelos Iliakis; Michael Moschos
Journal:  Clin Ophthalmol       Date:  2009-06-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.