Interinstrument variability in Hertel-type exophthalmometers.

Tamsin J Sleep1, Ruth M Manners.   

Abstract

PURPOSE: To investigate variability in measurements of axial globe position taken with different exophthalmometers of the same style and manufacturer by the same observer.
METHODS: In this small clinical study, 10 Hertel-type mirror exophthalmometers were used to measure the degree of exophthalmos of a plastic head. The deepest points on the lateral orbital margins were marked to ensure accurate placement of the instrument. A single observer, masked to previous measurements, recorded multiple measurements with each instrument over several days.
RESULTS: Variation was seen between instruments from different manufacturers and more interestingly, between the same design of instruments from the same company. The measurement of axial globe position varied by up to 2.9 mm between instruments of the same make. There was also variation in the measurement of the relative axial globe position of the model. This measurement varied from 0.6 to 2.4 mm. The mean variation of readings taken with the same instrument was 0.5 mm.
CONCLUSIONS: It is already considered good practice to minimize interobserver error by recording the make of exophthalmometer used. This small study also suggests that the same instrument should be used for each examination to minimize error caused by variation between instruments of the same make. Variation of 1.5 to 2.0 mm could affect patient treatment when a difference in measurement of 2 mm is regarded as significant during serial or relative exophthalmometry.

Entities:  

Mesh:

Year:  2002        PMID: 12142756     DOI: 10.1097/00002341-200207000-00004

Source DB:  PubMed          Journal:  Ophthalmic Plast Reconstr Surg        ISSN: 0740-9303            Impact factor:   1.746


  13 in total

1.  Slit-lamp exophthalmometry, a novel technique.

Authors:  Yehoshua Almog; Eli Rosen; Arie Y Nemet
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-04-01       Impact factor: 3.117

2.  Clinical features of thyroid-associated ophthalmopathy in clinically euthyroid Korean patients.

Authors:  S Y Jang; S Y Lee; E J Lee; J S Yoon
Journal:  Eye (Lond)       Date:  2012-06-29       Impact factor: 3.775

3.  Axial Globe Position Measurement: A Prospective Multicenter Study by the International Thyroid Eye Disease Society.

Authors:  Chad M Bingham; Jennifer A Sivak-Callcott; Matthew J Gurka; John Nguyen; Jeffery P Hogg; Steve E Feldon; Aaron Fay; Lay-Leng Seah; Diego Strianese; Vikram D Durairaj; Jimmy Uddin; Martin H Devoto; Matheson Harris; Justin Saunders; Tammy H Osaki; Audrey Looi; Livia Teo; Brett W Davies; Andrea Elefante; Sunny Shen; Tony Realini; William Fischer; Michael Kazim
Journal:  Ophthalmic Plast Reconstr Surg       Date:  2016 Mar-Apr       Impact factor: 1.746

4.  Validation of Modified Hertel Exophthalmometer.

Authors:  Rizwana Fathima Jamal; Emmanuel Azariah; Deepak Pandyan; Ravindran Chinnaswami
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2020-09-11

5.  Normal values of Hertel exophthalmometry in a Chinese Han population from Shenyang, Northeast China.

Authors:  Dan Wu; Xin Liu; Di Wu; Xin Di; Haixia Guan; Zhongyan Shan; Weiping Teng
Journal:  Sci Rep       Date:  2015-02-23       Impact factor: 4.379

6.  Association of anthropometric markers with globe position: A population-based MRI study.

Authors:  Patrick Schmidt; Robert Kempin; Sönke Langner; Achim Beule; Stefan Kindler; Thomas Koppe; Henry Völzke; Till Ittermann; Clemens Jürgens; Frank Tost
Journal:  PLoS One       Date:  2019-02-07       Impact factor: 3.240

7.  Hertel exophthalmometer versus computed tomography scan in proptosis estimation in thyroid-associated orbitopathy.

Authors:  Na Ri Park; Jun Hyung Moon; Jeong Kyu Lee
Journal:  Clin Ophthalmol       Date:  2019-08-02

8.  Management of Orbital Masses: Outcomes of Endoscopic and Combined Approaches With No Orbital Reconstruction.

Authors:  Ryan A Rimmer; Alexander E Graf; Judd H Fastenberg; Jurij Bilyk; Gurston G Nyquist; Marc R Rosen; Michael P Rabinowitz; Mindy R Rabinowitz
Journal:  Allergy Rhinol (Providence)       Date:  2020-01-14

9.  Measurement of proptosis using computed tomography based three-dimensional reconstruction software in patients with Graves' orbitopathy.

Authors:  Jung Huh; Sang Joon Park; Jeong Kyu Lee
Journal:  Sci Rep       Date:  2020-09-03       Impact factor: 4.379

10.  The feasibility analysis of calculating proptosis by simple Heron's formula.

Authors:  Weili Zhang; Qinying Huang; Yile Chen; Jinying Li
Journal:  PLoS One       Date:  2020-06-01       Impact factor: 3.752

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.