Literature DB >> 34471472

Validation of Modified Hertel Exophthalmometer.

Rizwana Fathima Jamal1, Emmanuel Azariah1, Deepak Pandyan1, Ravindran Chinnaswami1.   

Abstract

STUDY
DESIGN: This is a correlation study.
OBJECTIVE: The objective was to obtain normative values from modified Hertel exophthalmometer, to compare difference in values between modified and standard Hertel exophthalmometer and to statistically analyze for correlation.
METHODS: The study to validate the modification done to Hertel exophthalmometer was performed at the Department of Oral and Maxillofacial Surgery at our institute. Informed consent was obtained from patients; measurement in 100 volunteers was made with Hertel and modified Hertel exophthalmometer by a single observer, and the measurements taken by the 2 devices were tabulated along with variables such as age and gender. Data were collected and statistical analysis was performed to compare and correlate the values between standard and modified exophthalmometer. Differences of ocular protrusion values in gender and age were also calculated.
RESULTS: There has been no statistically significant difference in the normative data between standard Hertel and modified Hertel exophthalmometer and in relation to age and gender.
CONCLUSIONS: The modification of the Hertel exophthalmometer is validated and can also be used for relative exophthalmometry to assess the need for treatment and to evaluate the globe position and orbital volume intraoperatively. The proposed modification of the instrument allows it to be used in cases when one of the lateral orbital walls is fractured or displaced due to trauma.
© The Author(s) 2020.

Entities:  

Keywords:  Hertel exophthalmometer; exophthalmometry

Year:  2020        PMID: 34471472      PMCID: PMC8385629          DOI: 10.1177/1943387520954625

Source DB:  PubMed          Journal:  Craniomaxillofac Trauma Reconstr        ISSN: 1943-3875


  38 in total

1.  Relationship of the eye to the bony orbit, with clinical correlations.

Authors:  R A Goldberg; A Relan; J Hoenig
Journal:  Aust N Z J Ophthalmol       Date:  1999-12

2.  Errors of single-mirror or prism Hertel exophthalmometers and recommendations for minimizing the errors.

Authors:  W Terry Frueh; Bartley R Frueh
Journal:  Ophthalmic Plast Reconstr Surg       Date:  2007 May-Jun       Impact factor: 1.746

3.  OPHTHALMOLOGICAL SOCIETY-ANNUAL CONGRESS.

Authors: 
Journal:  Br J Ophthalmol       Date:  1938-06       Impact factor: 4.638

4.  Comparison of unilateral and simultaneous bilateral measurement of the globe position, using the Hertel exophthalmometer.

Authors:  Hossein Ameri; Sinead Fenton
Journal:  Ophthalmic Plast Reconstr Surg       Date:  2004-11       Impact factor: 1.746

5.  When is enophthalmos "significant"?

Authors:  Lily Koo; Mark P Hatton; Peter A D Rubin
Journal:  Ophthalmic Plast Reconstr Surg       Date:  2006 Jul-Aug       Impact factor: 1.746

6.  Dimensions and volumes of the orbit and orbital fat in posttraumatic enophthalmos.

Authors:  G Ramieri; M C Spada; S D Bianchi; S Berrone
Journal:  Dentomaxillofac Radiol       Date:  2000-09       Impact factor: 2.419

7.  Relationship between the extent of fracture and the degree of enophthalmos in isolated blowout fractures of the medial orbital wall.

Authors:  H R Jin; S O Shin; M J Choo; Y S Choi
Journal:  J Oral Maxillofac Surg       Date:  2000-06       Impact factor: 1.895

8.  Determination of the normal range of exophthalmometric values for black and white adults.

Authors:  M E Migliori; G J Gladstone
Journal:  Am J Ophthalmol       Date:  1984-10-15       Impact factor: 5.258

9.  The Effect of Age, Gender, Refractive Status and Axial Length on the Measurements of Hertel Exophthalmometry.

Authors:  Omer Karti; Ozlem B Selver; Eyyup Karahan; Mehmet O Zengin; Murat Uyar
Journal:  Open Ophthalmol J       Date:  2015-05-29

10.  The Correlation between the Degree of Enophthalmos and the Extent of Fracture in Medial Orbital Wall Fracture Left Untreated for Over Six Months: A Retrospective Analysis of 81 Cases at a Single Institution.

Authors:  Yun Sik Sung; Chan Min Chung; In Pyo Hong
Journal:  Arch Plast Surg       Date:  2013-07-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.