Literature DB >> 12127607

Six-month angiographic and 12-month clinical follow-up of MultiLink long (25 to 35 mm) stents for long coronary narrowings in patients with angina pectoris.

John A Ormiston1, Mark W I Webster, Peter N Ruygrok, Ian T Meredith, Justin P Ardill, Christopher E Buller, Donald R Ricci, Charles Chan, Gerard P Devlin, James T Stewart, Ian M Penn, Sue Price, Bruce Webber, Teena West.   

Abstract

There are limited prospective angiographic data on stent deployment for long coronary lesions. This multicenter prospective study enrolled 120 patients with a single de novo stenosis >20 mm in length, in a native vessel > or =3 mm diameter, suitable for a MultiLink stent 25 to 35 mm in length with additional stent deployment if required. Quantitative angiography before and immediately after stenting and at 6-month follow-up assessed restenosis for the complete lesion and for 5-mm segments of the stented and adjacent nonstented vessel. By 1 year, myocardial infarction had occurred in 3% and target vessel repeat revascularization in 12% of patients. The mean stented length (35.8 +/- 14.6 mm) closely matched mean lesion length (30.1 +/- 13.5 mm). Restenosis to > or =50% diameter loss occurred in 32% of patients, but to > or =70% in only 8%. Of the 147 segments (5 mm in length) with baseline stenosis <25%, only 3 patients (2%) developed restenosis of > or =50%, and only in 1 of these was it > or =70%. Stenting of long narrowings is associated with good clinical outcome and a low rate of severe restenosis. Mildly diseased segments of long lesions covered by a stent rarely became severely narrowed and had negligible influence on the overall restenosis rate. These data support a strategy of full lesion coverage by stent deployment.

Entities:  

Mesh:

Year:  2002        PMID: 12127607     DOI: 10.1016/s0002-9149(02)02458-x

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  3 in total

1.  Cobalt-chromium stents in long lesions of large vessels: clinical and angiographic results.

Authors:  Iñigo Lozano; Carlos Cuellas; Pablo Avanzas; Armando Pérez de Prado; Concepción Suárez; Juan Rondan; Daehyun Lee; Jesus M de la Hera; Felipe Fernández; César Morís
Journal:  Tex Heart Inst J       Date:  2011

Review 2.  Coronary artery stents: identification and evaluation.

Authors:  J Butany; K Carmichael; S W Leong; M J Collins
Journal:  J Clin Pathol       Date:  2005-08       Impact factor: 3.411

Review 3.  Biocompatibility of Coronary Stents.

Authors:  Thamarasee M Jeewandara; Steven G Wise; Martin K C Ng
Journal:  Materials (Basel)       Date:  2014-01-28       Impact factor: 3.623

  3 in total