The measurement of linear accelerator isocenter motion using a three-micrometer device and an adjustable pointer.

J S Tsai1, B H Curran, E S Sternick, M J Engler.   

Abstract

PURPOSE: The small motions of the major axes of a linear accelerator observed during gantry and treatment table rotation were measured to improve beam-target alignment during stereotactic radiosurgery (SRS). METHODS AND MATERIALS: Measurements of gantry isocenter motion and table rotational axis wobble were performed with an adjustable front pointer and a three-micrometer device. Nominal gantry and table isocenters were specified. The gantry motion path and table isocenter coordinates were then applied to offset simulated treatment target coordinates so as to compensate for gantry sag. Target simulation films were examined to document improvement of beam-target alignment.
RESULTS: The overall precision of the measurement of gantry and table isocenter coordinates was 0.2 mm. Over gantry rotation of 0 to 360 degrees, the gantry isocenter was found to follow a pinched loop with a maximum point to point distance of 1 mm. Table axis motion was found to be negligible relative to the reproducibility of gantry isocenter motion. Thus, a table isocenter was defined that was invariant to table rotation.
CONCLUSION: Results indicate that the three-micrometer device and adjustable front pointer are useful tools for three-dimensional (3D) mapping of gantry, collimator and table isocenters and their motions. It is suggested that such measurements may be useful in the quality assurance of linear accelerators, particularly to improve beam-target alignment during SRS and other high dose external beam therapy.

Entities:  

Mesh:

Year:  1996        PMID: 12118550     DOI: 10.1016/0360-3016(95)02085-3

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  5 in total

1.  An isocenter detection and verification device for use in proton therapy.

Authors:  H Fuse; T Sakae; T Terunuma; M Sato; S Aoki
Journal:  Rev Sci Instrum       Date:  2012-12       Impact factor: 1.523

2.  A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications.

Authors:  Jacek M Chojnowski; Jonathan R Sykes; David I Thwaites
Journal:  J Appl Clin Med Phys       Date:  2021-05-30       Impact factor: 2.102

Review 3.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

4.  A quality assurance method with submillimeter accuracy for stereotactic linear accelerators.

Authors:  Jimm Grimm; Shu-Ya Lisa Grimm; Indra J Das; Yunping Zhu; Inhwan Yeo; Jinyu Xue; Larry Simpson; Dayee Jacob; Abhirup Sarkar
Journal:  J Appl Clin Med Phys       Date:  2010-12-08       Impact factor: 2.102

5.  A photographic technique for quick assessment of mechanical isocenter of a linear accelerator.

Authors:  Janne Heikkilä; Jan Seppälä; Jouni Nuutinen; Antti Vanne; Tapani Lahtinen
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2017-04-28
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.