Literature DB >> 12108660

Ion chromatographic determination of trace hydroxylamine in waste streams generated by a pharmaceutical reaction process.

Peter N Fernando1, Ignatius N Egwu, Mumtaz S Hussain.   

Abstract

Hydroxylamine is a key raw material used in a synthetic drug process at Pharmacia. Since hydroxylamine is harmful to microorganisms, concentrations above 5 ppm could interfere with the biological sewage plant performance. This necessitated the development of a sensitive analytical method for detecting low levels of hydroxylamine in the waste streams generated from the pharmaceutical process. The present report describes a cation-exchange chromatographic method coupled with pulsed amperometric detection at a gold electrode for trace analysis of hydroxylamine. This method was evaluated by generating data on the parameters of specificity, precision, linearity, recovery and sensitivity.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12108660     DOI: 10.1016/s0021-9673(02)00145-0