Loop memory of haptic materials in posterior chamber intraocular lenses.

Andrea M Izak1, Liliana Werner, David J Apple, Tamer A Macky, Rupal H Trivedi, Suresh K Pandey.   

Abstract

PURPOSE: To compare the shape recovery ratios after compression of haptic materials used in the manufacture of intraocular lenses (IOLs).
SETTING: Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA.
METHODS: The loop memory of 40 silicone-optic posterior chamber IOLs was studied. All the IOLs had modified-C haptics made of poly(methyl methacrylate) (PMMA; n = 10), polyimide (n = 10), polyvinylidene fluoride (PVDF; n = 10), and polypropylene (PP; n = 10). After the overall diameter of each lens was measured (day 0), the lenses were inserted into plastic wells (9.5 mm in diameter) and immersed in water (37 degrees C) for 1 month. They were then placed on an open plate and allowed to reexpand for 2 months. Overall diameter measurements were performed within 5 minutes of the IOLs' removal from the wells and at subsequent time points (days 14, 28, 30, 60, 74, 88, and 95).
RESULTS: The loop memory of each lens was expressed as the difference between the initial overall diameter measurement (pretest) and the measurement at each time point; the lower the value, the higher the memory. The overall difference among the 4 groups was statistically significant at each time point (P < or = .001). From days 30 to 95, silicone-PMMA, silicone-elastimide, and silicone-PVDF IOLs had similar loop memory mean values, which were significantly lower than the mean value of silicone-PP IOLs (P <.05). The latter design tended to be deformed after removal from the wells, with increased optic-haptic angulation.
CONCLUSION: Studying the loop memory of haptic materials (PMMA, polyimide, PVDF, and PP) used in the manufacture of posterior chamber IOLs can help surgeons choose an appropriate IOL for each patient.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12106733     DOI: 10.1016/s0886-3350(02)01326-3

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Light-adjustable lens.

Authors:  Daniel M Schwartz
Journal:  Trans Am Ophthalmol Soc       Date:  2003

2.  Evaluation of the anterior segment parameters after Nd: YAG laser Capsulotomy: Effect the design of intraocular lens Haptic.

Authors:  Berkay Akmaz; Fahrettin Akay
Journal:  Pak J Med Sci       Date:  2018 Mar-Apr       Impact factor: 1.088

Review 3.  Posterior Capsule Opacification after Cataract Surgery via Implantation with Hydrophobic Acrylic Lens Compared with Silicone Intraocular Lens: A Systematic Review and Meta-Analysis.

Authors:  Ye Rin Kwon; Yoo Na Hwang; Sung Min Kim
Journal:  J Ophthalmol       Date:  2022-02-25       Impact factor: 1.909

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.