| Literature DB >> 12106506 |
Tina Mygind1, Svend Birkelund, Niels H Birkebaek, Lars Østergaard, Jørgen Skov Jensen, Gunna Christiansen.
Abstract
BACKGROUND: Chlamydia pneumoniae infection has been detected by serological methods, but PCR is gaining more interest. A number of different PCR assays have been developed and some are used in combination with serology for diagnosis. Real-time PCR could be an attractive new PCR method; therefore it must be evaluated and compared to conventional PCR methods.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12106506 PMCID: PMC117782 DOI: 10.1186/1471-2180-2-17
Source DB: PubMed Journal: BMC Microbiol ISSN: 1471-2180 Impact factor: 3.605
Sample volumes used in the two PCR methods
| Method 1 | Method 2 | |
| Standard samples | 2 μl | - |
| Reference samples | 6.8 μl | 10 μl |
| Clinical samples | 2 μl | 10 μl |
Reference samples analysed with the two PCR assays. * The unit of the numbers is copies/μl and the numbers are the mean of two measurements from two different runs. # The dilution series is serial and four-fold. Dilution no. 1 has the highest concentration and dilution no. 10 has the lowest concentration. Method 1 is the real-time PCR and method 2 is the conventional PCR.
| 1 (high) | 76.60 | + |
| 2 | 7.86 | + |
| 3 | 2.35 | + |
| 4 | 0.63 | + |
| 5 | 0.05 | + |
| 6 | 0.00 | - |
| 7 | 0.00 | - |
| 8 | 0.00 | - |
| 9 | 0.00 | - |
| 10 (low) | 0.00 | - |
| 11 (NK) | 0.00 | - |
Figure 1Determination of the efficiency for the standard, reference and clinical samples. The threshold cycle is the cycle number at which the fluorescence curve for a given sample crosses the user-defined noise band. Log to the concentration for the reference samples is set as log to arbitrary numbers with four-fold difference as the reference samples are four-fold serial dilutions. The actual concentration is not needed when determining the efficiency as it depends on the slope of the line only.
Figure 2Effect of difference in efficiency between standard samples and samples. This figure illustrates the error in determination of concentration by real-time PCR at cycle no. 30 when the efficiencies differ between standard samples and samples. Number of cycles: n, Efficiency standard samples: Estd, efficiency reference samples: Ers, Number of amplicons at cycle n: Nn, Number of amplicons at cycle zero: N0. When the two equations are divided by eachother it is seen that there is an 23-fold underestimation of the concentration when Estd = 2.0 and Ers = 1.8. From this calculation it is also evident that at higher cycle number the error in concentration determination is larger.
Positive samples found among 136 clinical samples analysed with PCR methods 1 and 2. *copies/μl is the mean of two measurements from two different runs.
| + | 9.73 | + | |
| + | 1.08 | + | |
| + | 750.80 | + |
Determination of PCR efficiency in 20 spiked clinical samples. S.E.: standard error, Eff.: Efficiency, std. standard samples. For each of 20 randomly selected negative samples spiked with C. pneumoniae DNA, the slope was determined as described and the efficiency calculated from the slope. The slope of each sample was compared to the slope of the standard from the same run with a t-test and the P-values are shown. Therefore, some of the samples are compared to the same standard.
| 1 | -3.064 | 0.215 | 2.12 | -3.406 | 0.232 | 1.97 | 0.28 |
| 60 | -3.190 | 0.273 | 2.06 | 0.55 | |||
| 14 | -2.873 | 0.147 | 2.23 | -3.266 | 0.264 | 2.02 | 0.19 |
| 31 | -3.222 | 0.091 | 2.04 | -3.623 | 0.340 | 1.89 | 0.25 |
| 32 | -3.549 | 0.156 | 1.91 | 0.84 | |||
| 78 | -3.328 | 0.153 | 2.00 | 0.43 | |||
| 84 | -4.318 | 0.512 | 1.70 | -3.321 | 0.180 | 2.00 | 0.066 |
| 90 | -3.379 | 0.177 | 1.98 | 0.82 | |||
| 97 | -3.503 | 0.291 | 1.93 | -3.716 | 0.187 | 1.86 | 0.54 |
| 101 | -3.266 | 0.085 | 2.02 | 0.028 | |||
| 103 | -3.388 | 0.091 | 1.97 | 0.12 | |||
| 118 | -3.398 | 0.189 | 1.97 | -3.492 | 0.094 | 1.93 | 0.65 |
| 127 | -3.495 | 0.145 | 1.93 | 0.99 | |||
| 131 | -2.809 | 0.108 | 2.27 | 0.00 | |||
| 138 | -3.545 | 0.183 | 1.91 | -3.300 | 0.101 | 2.01 | 0.24 |
| 145 | -3.396 | 0.134 | 1.97 | 0.28 | |||
| 150 | -3.652 | 0.394 | 1.88 | 0.39 | |||
| 159 | -3.491 | 0.133 | 1.93 | -3.615 | 0.056 | 1.89 | 0.39 |
| 160 | -2.135 | 0.313 | 2.94 | 0.00 | |||
| 171 | -3.531 | 0.088 | 1.92 | 0.42 |