Literature DB >> 12096272

Effect of endovascular stent strut geometry on vascular injury, myointimal hyperplasia, and restenosis.

Timothy M Sullivan1, Stephen D Ainsworth, Eugene M Langan, Spence Taylor, Bruce Snyder, David Cull, Jerry Youkey, Martine Laberge.   

Abstract

PURPOSE: Early restenosis and the development of myointimal hyperplasia in stented blood vessels have been attributed to deep vascular injury with fracture of the internal elastic lamina (IEL). The purpose of this study was the evaluation of the vascular wall response to superficial injury (without IEL rupture) after balloon angioplasty and intravascular stent placement in porcine arteries and the determination of the effect of stent strut geometry on the degree of vessel injury and early restenosis.
METHODS: Balloon-expandable stainless-steel stents were placed into the iliac arteries of 10 Sinclair miniature swine that had been fed an atherogenic diet. A Palmaz stent, with rectangular struts and smooth corners, was randomly assigned to one iliac artery (group 1), and a novel stent, which was designed and manufactured in the laboratory with thicker struts and sharper corners specifically to induce large wall stress concentrations, was placed in the contralateral iliac artery (group 2). Intravascular ultrasound scan was used in all deployments to ensure accurate balloon sizing and to avoid stent overexpansion and deep vascular injury. At 90 days after implantation, the animals were killed, the stented vessels harvested, and histomorphometric analysis performed.
RESULTS: Deployment of novel stents in group 2 resulted in a statistically higher incidence rate of deep vascular injury (fracture of the IEL) compared with group 1, despite identical balloon size used for deployment (with Student t test, P <.05). Vessels with deep injury showed a 10-fold increase in myointimal thickening compared with those vessels in which the IEL remained intact. A statistically higher restenosis rate was observed for group 2 (33.5% +/- 19.90%) compared with group 1 (20.39% +/- 14.70%). For both stent designs, there was a trend toward lower degrees of restenosis within the mid-portion of the stent. For superficially injured arteries in both groups, no correlation was observed between the amount of vessel wall/medial layer compression and the development of restenosis from myointimal hyperplasia.
CONCLUSION: Maintenance of an intact IEL is an important factor in the prevention of myointimal hyperplasia and restenosis in stented porcine iliac arteries. The alteration of stent strut height and geometry does not significantly affect restenosis and the development of myointimal hyperplasia in vessels with superficial injury. Superficial injury elicits a response that is independent of stent strut geometry and vessel wall compression. Stent strut profile may, however, increase local vessel wall stress concentrations, leading to IEL rupture and an exaggerated response to injury.

Entities:  

Mesh:

Year:  2002        PMID: 12096272     DOI: 10.1067/mva.2002.122878

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  9 in total

1.  In vivo Evaluation of Cenderitide-Eluting Stent (CES) II.

Authors:  Yingying Huang; Xu Wen Ng; Soon Ghim Lim; Horng Haur Chen; John C Burnett; Yin Chiang Freddy Boey; Subbu S Venkatraman
Journal:  Ann Biomed Eng       Date:  2015-07-16       Impact factor: 3.934

2.  Comparison of in vitro human endothelial cell response to self-expanding stent deployment in a straight and curved peripheral artery simulator.

Authors:  Ríona Ní Ghriallais; Laoise McNamara; Mark Bruzzi
Journal:  J R Soc Interface       Date:  2013-01-30       Impact factor: 4.118

3.  Flow-regulated endothelial glycocalyx determines metastatic cancer cell activity.

Authors:  Solomon A Mensah; Alina A Nersesyan; Ian C Harding; Claire I Lee; Xuefei Tan; Selina Banerjee; Mark Niedre; Vladimir P Torchilin; Eno E Ebong
Journal:  FASEB J       Date:  2020-03-13       Impact factor: 5.191

4.  Endovascular treatment of the vertebral artery origin stenosis by using the closed-cell, self-expandable Carotid Wallstent.

Authors:  Jun-Kyeung Ko; Chang-Hwa Choi; Lee Hwangbo; Hie-Bum Suh; Tae-Hong Lee; Han-Jin Cho; Sang-Min Sung
Journal:  Interv Neuroradiol       Date:  2020-06-20       Impact factor: 1.610

5.  Endovascular Stenting for Non-Traumatic Celiac Artery Stenosis Following Pancreatoduodenectomy.

Authors:  Si-Yuan Yao; Shintaro Yagi; Hiroyuki Ueda
Journal:  Ann Vasc Dis       Date:  2016-06-02

6.  Laparoscopic treatment of celiac axis compression syndrome (CACS) and hiatal hernia: Case report with bleeding complications and review.

Authors:  Lorenzo di Libero; Antonio Varricchio; Ernesto Tartaglia; Igino Iazzetta; Alberto Tartaglia; Antonella Bernardo; Rosanna Bernardo; Giovangiuseppe Triscino; Domenico Lo Conte
Journal:  Int J Surg Case Rep       Date:  2013-07-26

7.  Native extracellular matrix orientation determines multipotent vascular stem cell proliferation in response to cyclic uniaxial tensile strain and simulated stent indentation.

Authors:  P S Mathieu; E Fitzpatrick; M Di Luca; P A Cahill; C Lally
Journal:  Biochem Biophys Rep       Date:  2021-12-23

8.  Self-expandable nitinol stent placement in homocysteinemic porcine aorta.

Authors:  Luís Henrique Gil França; Adamastor Humberto Pereira; Sílvio César Perini
Journal:  Clinics (Sao Paulo)       Date:  2008-04       Impact factor: 2.365

9.  Procedural advantages of a novel coronary stent design with ultra-thin struts and bioabsorbable abluminal polymer coating in an all-comers registry.

Authors:  Haitham Abu Sharar; Bruna Gomes; Emmanuel Chorianopoulos; Ziya Kaya; Christian A Gleissner; Hugo A Katus; Raffi Bekeredjian
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-09-21       Impact factor: 1.426

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.