Literature DB >> 12074450

Two dental implants designed for immediate loading: a finite element analysis.

Laurent Pierrisnard1, Guy Hure, Michel Barquins, Daniel Chappard.   

Abstract

PURPOSE: The aim of this study was to evaluate by finite element analysis the influence of the design of 3 different dental implants on micromovements, cervical shearing stress intensity, and stress distribution after occlusal loading.
MATERIALS AND METHODS: The first investigated implant was a classical cylinder, the second was reinforced by 2 bicortical locking pins, and the third was an expanding dental implant. The parameters analyzed were the implant's geometry, the quality of the cancellous bone, and the orientation of occlusal loading.
RESULTS: It was found that initial stability of the locking pin implant was greater than the initial stability of the other investigated implant designs, regardless of the quality of cancellous bone and orientation of occlusal loading; in low-rigidity cancellous bone, under a horizontal load (500 N), decreasing displacement compared to those of the other investigated implants was 16 microm. The apical expansion and locking pin implants exhibited favorable behavior regarding the distribution and intensity of cervical shearing stresses; in low-rigidity cancellous bone, under horizontal load, decreasing cervical stresses compared with those of the cylindric implant were 10 MPa for the apical expansion implant and 150 MPa for the locking pin implant. DISCUSSION: For the cylindric implant, stresses were concentrated in the neck region; for the apical expansion implant, stresses were evenly distributed from the neck to the apex of the implant. For the locking pin implant, stresses around the neck were moderate and appeared concentrated around the pins.
CONCLUSIONS: Initial stability of the pin implant was greater than that of the expanding implant, but the expanding implant showed the most favorable stress distribution.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12074450

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  12 in total

1.  Stress distribution in implant retained finger prosthesis: a finite element study.

Authors:  Pokpong Amornvit; Dinesh Rokaya; Konrawee Keawcharoen; Nimit Thongpulsawasdi
Journal:  J Clin Diagn Res       Date:  2013-12-15

2.  In vivo bone strain and finite element modeling of a rhesus macaque mandible during mastication.

Authors:  Olga Panagiotopoulou; José Iriarte-Diaz; Simon Wilshin; Paul C Dechow; Andrea B Taylor; Hyab Mehari Abraha; Sharifah F Aljunid; Callum F Ross
Journal:  Zoology (Jena)       Date:  2017-09-01       Impact factor: 2.240

3.  Nonlinear finite element analysis of three implant- abutment interface designs.

Authors:  Chun-Bo Tang; Si-Yu Liul; Guo-Xing Zhou; Jin-Hua Yu; Guang-Dong Zhang; Yi-Dong Bao; Qiu-Ju Wang
Journal:  Int J Oral Sci       Date:  2012-06       Impact factor: 6.344

4.  Evaluation of Micromovements and Stresses around Single Wide-Diameter and Double Implants for Replacing Mandibular Molar: A Three-Dimensional FEA.

Authors:  Shrikar R Desai; I Karthikeyan; Rika Singh
Journal:  ISRN Dent       Date:  2012-03-07

5.  2D FEA of evaluation of micromovements and stresses at bone-implant interface in immediately loaded tapered implants in the posterior maxilla.

Authors:  Shrikar R Desai; Rika Singh; I Karthikeyan
Journal:  J Indian Soc Periodontol       Date:  2013-09

6.  Influence of three different implant thread designs on stress distribution: A three-dimensional finite element analysis.

Authors:  Mansi Manish Oswal; Ulhas N Amasi; Manish S Oswal; Ashish S Bhagat
Journal:  J Indian Prosthodont Soc       Date:  2016 Oct-Dec

7.  Stress distribution pattern of screw-retained restorations with segmented vs. non-segmented abutments: A finite element analysis.

Authors:  Shima Aalaei; Zahra Rajabi Naraki; Fatemeh Nematollahi; Elaheh Beyabanaki; Afsaneh Shahrokhi Rad
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-09-20

8.  Finite Element Analysis of Dental Implants with Zirconia Crown Restorations: Conventional Cement-Retained vs. Cementless Screw-Retained.

Authors:  Jae-Hyun Lee; Ho Yeol Jang; Su Young Lee
Journal:  Materials (Basel)       Date:  2021-05-19       Impact factor: 3.623

9.  3D finite element analysis of immediate loading of single wide versus double implants for replacing mandibular molar.

Authors:  Shrikar R Desai; I Karthikeyan; Reetika Gaddale
Journal:  J Indian Soc Periodontol       Date:  2013-11

10.  Biomechanical effects of original equipment manufacturer and aftermarket abutment screws in zirconia abutment on dental implant assembly.

Authors:  Yu-Ling Wu; Ming-Hsu Tsai; Hung-Shyong Chen; Yu-Tsen Chang; Tsai-Te Lin; Aaron Yu-Jen Wu
Journal:  Sci Rep       Date:  2020-10-27       Impact factor: 4.379