Literature DB >> 12066638

Alumina ceramic implant abutments used for single-tooth replacement: a prospective 1- to 3-year multicenter study.

B Andersson1, A Taylor, B R Lang, H Scheller, P Schärer, J A Sorensen, D Tarnow.   

Abstract

PURPOSE: The aim of the study was to compare results after 1 and 3 years when single crowns supported by CerAdapt (test) ceramic abutments or CeraOne (control) titanium abutments were loaded.
MATERIALS AND METHODS: The material was divided into two groups: in group A, 69 (34 test, 35 control) abutments/crowns from all involved clinics were followed for 1 year; and in group B, 20 (10 test, 10 control) abutments/crowns from one of the clinics were followed for 3 years.
RESULTS: No implant failed, giving a cumulative success rate of 100% for the implants. Two CerAdapt abutments in group A fractured, giving a cumulative success rate of 93% for the CerAdapt and 100% for the CeraOne abutments after 1 year. No abutment failed in group B, giving a cumulative success rate of 100% both for the CerAdapt and the CeraOne abutments between 1 and 3 years. The CerAdapt fractures might have been due to the fact that the abutments were impaired through too-extensive preparation and/or had been exposed to a too-high bending moment. Almost no marginal bone loss was recorded, indicating a stable bone situation both at CerAdapt and CeraOne abutments on single-tooth implants. Healthy conditions, with a relatively stable level of the periimplant mucosa in relation to the abutment/crown, were recorded for soft tissues both at CerAdapt and CeraOne abutments. Both clinicians and patients rated the esthetic result as excellent for practically all cases.
CONCLUSION: The results demonstrate the esthetic possibilities and the safety of single-tooth replacement when accepted treatment concepts are followed and documented components are used. The tested abutments worked well, although the fractured CerAdapt abutments indicate that ceramic abutments are more sensitive to handling procedures than the titanium abutments.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 12066638

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  10 in total

Review 1.  WITHDRAWN: Interventions for replacing missing teeth: partially absent dentition.

Authors:  Elliot Abt; Alan B Carr; Helen V Worthington
Journal:  Cochrane Database Syst Rev       Date:  2019-07-17

2.  Precision of fit: zirconia three-unit fixed dental prostheses.

Authors:  Florian Beuer; Michael Naumann; Wolfgang Gernet; John A Sorensen
Journal:  Clin Oral Investig       Date:  2008-09-04       Impact factor: 3.573

3.  Impact of abutment material on peri-implant soft tissue color. An in vitro study.

Authors:  Leticia Sala; Antonio Bascones-Martínez; Ana Carrillo-de-Albornoz
Journal:  Clin Oral Investig       Date:  2016-11-22       Impact factor: 3.573

4.  Assessment of human gingival fibroblast interaction with dental implant abutment materials.

Authors:  Vygandas Rutkunas; Virginija Bukelskiene; Vaidotas Sabaliauskas; Evaldas Balciunas; Mangirdas Malinauskas; Daiva Baltriukiene
Journal:  J Mater Sci Mater Med       Date:  2015-03-25       Impact factor: 3.896

5.  Reliability and fatigue failure modes of implant-supported aluminum-oxide fixed dental prostheses.

Authors:  Christian F J Stappert; Marta Baldassarri; Yu Zhang; Felix Hänssler; Elizabeth D Rekow
Journal:  Clin Oral Implants Res       Date:  2011-09-05       Impact factor: 5.977

6.  Interaction of titanium, zirconia and lithium disilicate with peri-implant soft tissue: study protocol for a randomized controlled trial.

Authors:  Katharina Kuhn; Heike Rudolph; Michael Graf; Matthias Moldan; Shaoxia Zhou; Martin Udart; Andrea Böhmler; Ralph G Luthardt
Journal:  Trials       Date:  2015-10-15       Impact factor: 2.279

7.  Color variation induced by abutments in the superior anterior maxilla: an in vitro study in the pig gingiva.

Authors:  Ramin Atash; Mohamed-Reda Boularbah; Cetik Sibel
Journal:  J Adv Prosthodont       Date:  2016-12-15       Impact factor: 1.904

8.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

9.  Effects of Liner-Bonding of Implant-Supported Glass-Ceramic Crown to Zirconia Abutment on Bond Strength and Fracture Resistance.

Authors:  Yong-Seok Jang; Sang-Hoon Oh; Won-Suck Oh; Min-Ho Lee; Jung-Jin Lee; Tae-Sung Bae
Journal:  Materials (Basel)       Date:  2019-08-30       Impact factor: 3.623

Review 10.  Effects of modified abutment characteristics on peri-implant soft tissue health: A systematic review and meta-analysis.

Authors:  Ignacio Sanz-Martín; Ignacio Sanz-Sánchez; Ana Carrillo de Albornoz; Elena Figuero; Mariano Sanz
Journal:  Clin Oral Implants Res       Date:  2017-10-26       Impact factor: 5.977

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.