Image distortion in MRI-based polymer gel dosimetry of gamma knife stereotactic radiosurgery systems.

Yoichi Watanabe1, Gerard M Perera, Robertus B Mooij.   

Abstract

We have studied effects of MR (magnetic resonance) image distortion on polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems. MR images of BANG polymer gel phantoms were acquired by using a Hahn spin-echo sequence and a fast 3D imaging GRASS sequence. Image artifacts were studied by varying the directions of frequency encoding and the receiver bandwidth. The phantoms were also CT (computed tomography) scanned. The studies showed that the measured dose distributions are shifted by 1.8+/-0.5 mm (2 pixels) in the frequency encoding direction. The magnitude of the shift is inversely proportional to the receiver bandwidth in agreement with theory. Comparison of MRI with CT showed the same image shift. We concluded that the discrepancy is caused by MR image distortion due to a difference in susceptibility effects between the phantom and the fiducial markers of the Leksell localization box.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12033575     DOI: 10.1118/1.1470204

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Sensitivity calibration procedures in optical-CT scanning of BANG 3 polymer gel dosimeters.

Authors:  Y Xu; Cheng-Shie Wuu; Marek J Maryanski
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

Review 2.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

3.  MAGAT gel and EBT2 film-based dosimetry for evaluating source plugging-based treatment plan in Gamma Knife stereotactic radiosurgery.

Authors:  Gopishankar Natanasabapathi; Vivekanandhan Subbiah; Shashank Sharad Kale; Goura Kishor Rath; S Senthilkumaran; Sanjay Thulkar; Vellaiyan Subramani; M A Laviraj; Raj Kishor Bisht; A K Mahapatra
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

4.  MRI-based polymer gel dosimetry for validating plans with multiple matrices in Gamma Knife stereotactic radiosurgery.

Authors:  N Gopishankar; Yoichi Watanabe; Vivekanandhan Subbiah
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

Review 5.  Three-dimensional radiation dosimetry using polymer gel and solid radiochromic polymer: From basics to clinical applications.

Authors:  Yoichi Watanabe; Leighton Warmington; N Gopishankar
Journal:  World J Radiol       Date:  2017-03-28

6.  Evaluation of dose delivery accuracy of Gamma Knife by polymer gel dosimetry.

Authors:  Yoichi Watanabe; Tomohide Akimitsu; Yutaka Hirokawa; Rob B Mooij; G Mark Perera
Journal:  J Appl Clin Med Phys       Date:  2005-08-12       Impact factor: 2.102

Review 7.  Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D.

Authors:  Yves De Deene
Journal:  Gels       Date:  2022-09-19

8.  Image fusion pitfalls for cranial radiosurgery.

Authors:  Benjamin P Jonker
Journal:  Surg Neurol Int       Date:  2013-04-17
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.