Mohammed Abrar1, Paul D Martin.
Abstract
The validation of a method for the determination of mevalonic acid (MVA; after conversion to the lactone, MVAL) in human plasma, using high-performance liquid chromatography with tandem mass spectrometry (HPLC-MS-MS), is reported. MVAL and deuterated internal standard were extracted from human plasma samples using automated solid-phase extraction. Analysis was conducted by column-switching, reversed-phase LC-MS-MS, using two hyper-cross-linked styrene-divinylbenzene copolymer sorbent reversed-phase columns. An assay range of 0.2-35 ng/ml and a lower limit of quantitation (LLOQ) of 0.2 ng/ml were achieved with acceptable accuracy and precision. MVA was stable in plasma under a variety of storage conditions.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12031835 DOI: 10.1016/s1570-0232(02)00131-9