STUDY
OBJECTIVE: To compare cardiac output (CO) as measured by the arterial thermodilution technique using only a central venous catheter and an arterial catheter inserted into the axillary artery, with conventional CO measurement with thermodilution using a pulmonary artery (PA) catheter (PAC).
DESIGN: Prospective clinical study in which each patient served as his/her own control.
SETTING: General ICU of a large tertiary-care teaching hospital. PATIENTS: 22 patients who required invasive hemodynamic monitoring in the ICU. INTERVENTIONS AND MEASUREMENTS: CO measurements made using the PAC (COpa) were compared to bolus arterial thermodilution measurements (COax). The significance of acute changes in the continuous CO measurements during acute hemodynamic episodes was observed. MAIN
RESULTS: The correlation between the two techniques (COpa and COax) was R(2) = 0.82. There was a tendency for 5% overestimation of COpa by the COax. The SEM% (SEM/average CO) for COax and COpa was 2.6% and 3.2%, respectively. The bias between measurements was 0.27 +/- 0.67 L/min, and the limits of agreement (mean difference +/- 2 SD) from minus 1.07 L/min to 1.63 L/min.
CONCLUSIONS: In critically ill patients, in whom the measurement of CO is required, arterial thermodilution, using a central vein and the axillary artery is accurate and reproducible.

Entities:  

Mesh:

Year:  2002        PMID: 12031755     DOI: 10.1016/s0952-8180(02)00345-8

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  7 in total

Review 1.  Reproducibility of transpulmonary thermodilution cardiac output measurements in clinical practice: a systematic review.

Authors:  Raphaël Giraud; Nils Siegenthaler; Paolo Merlani; Karim Bendjelid
Journal:  J Clin Monit Comput       Date:  2016-01-11       Impact factor: 2.502

Review 2.  The transpulmonary thermodilution technique.

Authors:  Samir G Sakka; Daniel A Reuter; Azriel Perel
Journal:  J Clin Monit Comput       Date:  2012-07-18       Impact factor: 2.502

Review 3.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

4.  Refinement of pig retroperfusion technique: Global retroperfusion with ligation of the azygos connection preserves hemodynamic function in an acute infarction model in pigs (Sus scrofa domestica).

Authors:  Frank Harig; Evelyn Hoyer; Dirk Labahn; Joachim Schmidt; Michael Weyand; Stephan M Ensminger
Journal:  Comp Med       Date:  2010-02       Impact factor: 0.982

5.  Precision of the transpulmonary thermodilution measurements.

Authors:  Xavier Monnet; Romain Persichini; Mariem Ktari; Mathieu Jozwiak; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2011-08-27       Impact factor: 9.097

6.  Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patients.

Authors:  John F Stover; Reto Stocker; Renato Lenherr; Thomas A Neff; Silvia R Cottini; Bernhard Zoller; Markus Béchir
Journal:  BMC Anesthesiol       Date:  2009-10-12       Impact factor: 2.217

7.  Use of the PiCCO system in critically ill patients with septic shock and acute respiratory distress syndrome: a study protocol for a randomized controlled trial.

Authors:  Zhongheng Zhang; Xiao Xu; Min Yao; Huilan Chen; Hongying Ni; Haozhe Fan
Journal:  Trials       Date:  2013-02-01       Impact factor: 2.279

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.