Setting performance goals and evaluating total analytical error for diagnostic assays.

Jan S Krouwer1.   

Abstract

BACKGROUND: Total analytical error has been a useful metric both to assess laboratory assay quality and to set goals. It is often estimated by combining imprecision (SD) and average bias in the equation: total analytical error = bias + 1.65 x imprecision. This indirect estimation model (referred to as the simple combination model) leads to different estimates of total analytical error than that of a direct estimation method (referred to as the distribution-of-differences method) or of simulation.
METHODS: A review of the literature was undertaken to reconcile the different estimation approaches.
RESULTS: The simple combination model can underestimate total analytical error by neglecting random interference bias and by not properly treating other error sources such as linear drift and outliers. A simulation method to estimate total analytical error is outlined, based on the estimation and combination of total analytical error source distributions. Goals for each total analytical error source can be established by allocation of the total analytical error goal. Typically, the allocation is cost-based and uses the probability of combinations of error sources. The distribution-of-differences method, simple combination model, and simulation method to evaluate total analytical error are compared. Outlier results can profoundly influence quality, but their rates are seldom reported.
CONCLUSIONS: Total analytical error should be estimated either directly by the distribution-of-differences method or by simulation. A systems engineering approach that uses allocation of the total analytical error goal into error source goals provides a cost-effective approach to meeting total analytical error. Because outliers can cause serious laboratory error, the inclusion of outlier rate estimates from large studies (e.g., those conducted by manufacturers) would be helpful in assessing assay quality.

Entities:  

Mesh:

Year:  2002        PMID: 12029009

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  8 in total

Review 1.  Assessing the analytical performance of systems for self-monitoring of blood glucose: concepts of performance evaluation and definition of metrological key terms.

Authors:  Oliver Schnell; Rolf Hinzmann; Bernd Kulzer; Guido Freckmann; Michael Erbach; Volker Lodwig; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

Review 2.  Uncertainty in measurement: a review of monte carlo simulation using microsoft excel for the calculation of uncertainties through functional relationships, including uncertainties in empirically derived constants.

Authors:  Ian Farrance; Robert Frenkel
Journal:  Clin Biochem Rev       Date:  2014-02

3.  The chronic injury glucose error grid: a tool to reduce diabetes complications.

Authors:  Jan S Krouwer; George S Cembrowski
Journal:  J Diabetes Sci Technol       Date:  2014-10-14

4.  Evaluation of Sigma Metrics and Westgard Rule Selection and Implementation of Internal Quality Control in Clinical Chemistry Reference Laboratory, Ethiopian Public Health Institute.

Authors:  Zeleke Geto; Tigist Getahun; Tadesse Lejisa; Yosef Tolcha; Demirew Bikila; Chala Bashea; Mehari Meles; Wossene Habtu; Genet Ashebir; Bikila Negasa; Meron Sileshi; Yeabkal Daniel; Andargachew Gashu; Feyissa Challa
Journal:  Indian J Clin Biochem       Date:  2021-08-02

5.  Evaluation of Sigma-Metric and Application of Quality Tools in Clinical Laboratory of a Tertiary Care Hospital.

Authors:  Puja Kumari Jha; Nirupama Sharma; Juhee Chandra; Rachna Agarwal
Journal:  Indian J Clin Biochem       Date:  2020-09-19

6.  Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer.

Authors:  Sarah Hindenberg; Stefanie Klenner-Gastreich; Nicole Kneier; Sabine Zielinsky; Kris Gommeren; Natali Bauer; Andreas Moritz
Journal:  BMC Vet Res       Date:  2017-05-30       Impact factor: 2.741

7.  Robust automated reading of the skin prick test via 3D imaging and parametric surface fitting.

Authors:  Jesus Pineda; Raul Vargas; Lenny A Romero; Javier Marrugo; Jaime Meneses; Andres G Marrugo
Journal:  PLoS One       Date:  2019-10-21       Impact factor: 3.240

Review 8.  Biological variation: Understanding why it is so important?

Authors:  Tony Badrick
Journal:  Pract Lab Med       Date:  2021-01-04
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.