Continuous ambulatory right heart pressure measurements with an implantable hemodynamic monitor: a multicenter, 12-month follow-up study of patients with chronic heart failure.

Anthony Magalski1, Philip Adamson, Frederick Gadler, Michael Böehm, David Steinhaus, Dwight Reynolds, Kathryn Vlach, Cecilia Linde, Bodo Cremers, Brandon Sparks, Tom Bennett.   

Abstract

BACKGROUND: We describe the performance of an implantable hemodynamic monitor (IHM) that allows continuous recording of heart rate, patient activity levels, and right ventricular systolic, right ventricular diastolic, and estimated pulmonary artery diastolic pressures. Pressure parameters derived from the implantable monitor were correlated to measurements made with a balloon-tipped catheter to establish accuracy and reproducibility over time in patients with chronic heart failure (CHF). METHODS AND
RESULTS: IHM devices were implanted in 32 patients with CHF (left ventricular ejection fraction, 29% +/- 11%; range, 14%-62%) and were tested with right heart catheterization at implantation and 3, 6, and 12 months later. Hemodynamic variables were digitally recorded simultaneously from the IHM and catheter. Values were recorded during supine rest, peak response of Valsalva maneuver, sitting, peak of a 2-stage (25-50 W) bicycle exercise test, and final rest period. The median of 21 paired beat-to-beat cardiac cycles was analyzed for each intervention. A total of 217 paired data values from all maneuvers were analyzed for 32 patients at implantation and 129 paired data values for 20 patients at 1 year. The IHM and catheter values were not different at baseline or at 1 year (P >.05). Combining all interventions, correlation coefficients were 0.96 and 0.94 for right ventricular systolic pressure, 0.96 and 0.83 for right ventricular diastolic pressure, and 0.87 and 0.87 for estimated pulmonary artery diastolic pressure at implantation and 1 year, respectively.
CONCLUSIONS: The IHM and a standard reference pressure system recorded comparable right heart pressure values in patients with CHF. This implantable pressure transducer is accurate over time and provides a means to precisely monitor the hemodynamic condition of patients with CHF in a continuous fashion.

Entities:  

Mesh:

Year:  2002        PMID: 12016628     DOI: 10.1054/jcaf.2002.32373

Source DB:  PubMed          Journal:  J Card Fail        ISSN: 1071-9164            Impact factor:   5.712


  30 in total

Review 1.  Nonpharmacologic options for the management of heart failure.

Authors:  Robin J Trupp; William T Abraham
Journal:  Curr Cardiol Rep       Date:  2003-05       Impact factor: 2.931

Review 2.  Continuous and less invasive central hemodynamic monitoring by blood pressure waveform analysis.

Authors:  Ramakrishna Mukkamala; Da Xu
Journal:  Am J Physiol Heart Circ Physiol       Date:  2010-07-09       Impact factor: 4.733

3.  Left ventricular epicardial admittance measurement for detection of acute LV dilation.

Authors:  John E Porterfield; Erik R Larson; James T Jenkins; Daniel Escobedo; Jonathan W Valvano; John A Pearce; Marc D Feldman
Journal:  J Appl Physiol (1985)       Date:  2010-12-09

Review 4.  Implantable hemodynamic monitoring devices in heart failure.

Authors:  Salpy V Pamboukian; Melissa C Smallfield; Robert C Bourge
Journal:  Curr Cardiol Rep       Date:  2006-05       Impact factor: 2.931

Review 5.  Microelectromechanical systems and nephrology: the next frontier in renal replacement technology.

Authors:  Steven Kim; Shuvo Roy
Journal:  Adv Chronic Kidney Dis       Date:  2013-11       Impact factor: 3.620

Review 6.  Pathophysiology of the transition from chronic compensated and acute decompensated heart failure: new insights from continuous monitoring devices.

Authors:  Philip B Adamson
Journal:  Curr Heart Fail Rep       Date:  2009-12

Review 7.  Implantable Sensors for Regenerative Medicine.

Authors:  Brett S Klosterhoff; Melissa Tsang; Didi She; Keat Ghee Ong; Mark G Allen; Nick J Willett; Robert E Guldberg
Journal:  J Biomech Eng       Date:  2017-02-01       Impact factor: 2.097

Review 8.  An updated review of cardiac devices in heart failure.

Authors:  C Murphy; H Zafar; F Sharif
Journal:  Ir J Med Sci       Date:  2017-03-25       Impact factor: 1.568

9.  Therapeutic implications of implantable device-based monitoring of patients with heart failure.

Authors:  Eric Popjes; John P Boehmer
Journal:  Curr Treat Options Cardiovasc Med       Date:  2008-09

10.  Monitoring of heart failure: comparison of left atrial pressure with intrathoracic impedance and natriuretic peptide measurements in an experimental model of ovine heart failure.

Authors:  Miriam T Rademaker; Christopher J Charles; Iain C Melton; A Mark Richards; Christopher M Frampton; Jeff Siou; Fujian Qu; Neal L Eigler; Dan Gutfinger; Richard W Troughton
Journal:  Clin Sci (Lond)       Date:  2011-03       Impact factor: 6.124

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.