Literature DB >> 12002210

The mechanical effects of suture anchor insertion angle for rotator cuff repair.

Frank A Liporace1, Christopher M Bono, Steven A Caruso, Bradley Weiner, Keith Penny, Andrew J Feldman, Mark G Grossman, Thomas R Haher.   

Abstract

Twelve matched pairs of humerii were instrumented with suture anchor at 90 degrees, 75 degrees, 45 degrees, and 30 degrees relative to the cortical border at the junction of the greater tuberosity and articular surface. Two fixtures were inserted into each specimen at different angles and loaded to failure. Suture anchors failed at an average of 171, 219, 169, and 192 N with 90 degrees, 75 degrees, 45 degrees, and 30 degrees insertion angles, respectively. No statistical difference was detected between groups (P=.08). Although previous authors have prescribed angles < or =45 degrees to improve pull-out strength, the current in vitro data does not support these recommendations.

Entities:  

Mesh:

Year:  2002        PMID: 12002210     DOI: 10.3928/0147-7447-20020401-14

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  6 in total

1.  Stress distribution inside bone after suture anchor insertion: simulation using a three-dimensional finite element method.

Authors:  Hirotaka Sano; Atsushi Takahashi; Daisuke Chiba; Taku Hatta; Nobuyuki Yamamoto; Eiji Itoi
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-05-24       Impact factor: 4.342

Review 2.  Rotator cuff tears: pathology and repair.

Authors:  Hemang Yadav; Shane Nho; Anthony Romeo; John D MacGillivray
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-12-23       Impact factor: 4.342

Review 3.  [Complications related to implants in arthroscopic shoulder surgery].

Authors:  O Lorbach; P Wilmes; P Brogard; R Seil
Journal:  Orthopade       Date:  2008-11       Impact factor: 1.087

4.  Medialization of medial row anchor via the Nevasier portal yield enhanced footprint and outcomes in medium-to-large rotator cuff tears.

Authors:  Seung-Bae Oh; Jae-Jung Jeong; Jong-Hun Ji; Kaushal Patel; Won-Ha Hwang; Joon-Hyung Cho
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-08-02       Impact factor: 4.342

5.  Effect of anchor threads on the pullout strength: A biomechanical study.

Authors:  Hideaki Nagamoto; Nobuyuki Yamamoto; Eiji Itoi
Journal:  J Orthop       Date:  2018-08-16

6.  Conventional rotator cuff versus all-suture anchors-A biomechanical study focusing on the insertion angle in an unlimited cyclic model.

Authors:  Dimitris Ntalos; Kay Sellenschloh; Gerd Huber; Daniel Briem; Klaus Püschel; Michael M Morlock; Karl-Heinz Frosch; Florian Fensky; Till Orla Klatte
Journal:  PLoS One       Date:  2019-11-27       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.