Ali S Meigooni1, Maung M Yoe-Sein, Awni Y Al-Otoom, Keith T Sowards.
Abstract
The TG-43 recommended dosimetric characteristics of a new 125I brachytherapy source have been experimentally and theoretically determined. The measurements were performed in Solid Water using LiF TLDs. The calculations were performed using Monte Carlo simulations in Solid Water and water. The measured data were compared with calculated values as well as the reported data in literature for other source designs. The dose rate constant this source in water was 1.01 +/- 3% cGy h(-1) U(-1) and the anisotropy constant was 0.956.Entities:
Mesh:
Substances:
Year: 2002 PMID: 11999157 DOI: 10.1016/s0969-8043(01)00258-5
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513