A 28-year-old man with patent foramen ovale and a prominent Eustachian valve with a history of transient ischemic attack underwent transcatheter closure using Amplatzer patent foramen ovale occluder. During deployment, some of the prominent valve tissue was entrapped on the delivery cable and a piece of the valve was extracted unintentionally. Anatomico-clinical and technical implications are discussed. Copyright 2002 Wiley-Liss, Inc.Entities: