Performance of latex and nonlatex medical examination gloves during simulated use.

Denise M Korniewicz1, Maher El-Masri, John M Broyles, Christopher D Martin, Kevin P O'connell.   

Abstract

BACKGROUND: In response to the rise in latex allergies, gloves made from a variety of nonlatex materials have been introduced into the health care environment. To date, at least 1 study, by Rego and Roley (1999), has reported that both latex and nitrile medical examination gloves provide comparable barrier protective qualities. The purpose of our study was to determine the effects of glove stress, type of material (vinyl, nitrile, copolymer, latex), and manufacturer on the barrier effectiveness of medical examination gloves.
METHOD: A total of 5510 medical examination gloves (1464 nitrile, 1052 latex, 1006 copolymer, and 1988 vinyl) were divided into 2 groups: stressed and unstressed. Unstressed gloves were visually inspected and water-tested according to the Food and Drug Administration water-testing standards. Stressed gloves were manipulated according to a designated stress protocol, visually inspected, and then subjected to the same Food and Drug Administration water-testing standards.
RESULTS: Our limited sample size demonstrated that nitrile gloves had the lowest failure rate (1.3%), followed by latex (2.2%); vinyl and copolymer gloves had the highest failure rate (both 8.2%). With use of a logistic regression analysis adjusting for manufacturer and stress, latex examination gloves were found to be 3 times more likely to fail than nitrile gloves (odds ratio, 3.2; 95% CI, 1.37-7.50). Nitrile gloves were also found to fail significantly less often than vinyl or copolymer gloves (odds ratio, 12.60; 95% CI, 5.80-27.40).
CONCLUSIONS: Nitrile examination gloves are a suitable alternative to latex, whereas vinyl and copolymer examination gloves were found to be less effective barriers. Further research is indicated to determine whether nitrile gloves can provide effective barrier qualities during clinical use versus laboratory simulations.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11944004     DOI: 10.1067/mic.2002.119512

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  10 in total

1.  2007 Guideline for Isolation Precautions: Preventing Transmission of Infectious Agents in Health Care Settings.

Authors:  Jane D Siegel; Emily Rhinehart; Marguerite Jackson; Linda Chiarello
Journal:  Am J Infect Control       Date:  2007-12       Impact factor: 2.918

2.  Guidelines for preventing infectious complications among hematopoietic cell transplantation recipients: a global perspective.

Authors:  Marcie Tomblyn; Tom Chiller; Hermann Einsele; Ronald Gress; Kent Sepkowitz; Jan Storek; John R Wingard; Jo-Anne H Young; Michael J Boeckh; Michael A Boeckh
Journal:  Biol Blood Marrow Transplant       Date:  2009-10       Impact factor: 5.742

3.  epic3: national evidence-based guidelines for preventing healthcare-associated infections in NHS hospitals in England.

Authors:  H P Loveday; J A Wilson; R J Pratt; M Golsorkhi; A Tingle; A Bak; J Browne; J Prieto; M Wilcox
Journal:  J Hosp Infect       Date:  2014-01       Impact factor: 3.926

4.  Tensile properties and integrity of clean room and low-modulus disposable nitrile gloves: a comparison of two dissimilar glove types.

Authors:  Robert N Phalen; Weng Kee Wong
Journal:  Ann Occup Hyg       Date:  2011-12-26

5.  epic2: National evidence-based guidelines for preventing healthcare-associated infections in NHS hospitals in England.

Authors:  R J Pratt; C M Pellowe; J A Wilson; H P Loveday; P J Harper; S R L J Jones; C McDougall; M H Wilcox
Journal:  J Hosp Infect       Date:  2007-02       Impact factor: 3.926

6.  Integrity of disposable nitrile exam gloves exposed to simulated movement.

Authors:  Robert N Phalen; Weng Kee Wong
Journal:  J Occup Environ Hyg       Date:  2011-05       Impact factor: 2.155

7.  Do Gloves Provide Adequate Protection Against Infection?

Authors:  Dylan Singh; Alan Parsa; Fereydoun D Parsa
Journal:  Aesthetic Plast Surg       Date:  2021-09-07       Impact factor: 2.326

8.  Recommended practices for prevention of transmissible infections in the perioperative practice setting.

Authors: 
Journal:  AORN J       Date:  2007-02       Impact factor: 0.676

9.  Effect of multiple alcohol-based hand rub applications on the tensile properties of thirteen brands of medical exam nitrile and latex gloves.

Authors:  Pengfei Gao; Matthew Horvatin; George Niezgoda; Robyn Weible; Ronald Shaffer
Journal:  J Occup Environ Hyg       Date:  2016-12       Impact factor: 2.155

10.  Dental professionals' knowledge and behavior towards utilization of gloves: a cross-sectional survey.

Authors:  Arun Suresh Dodamani; Harish Chaitram Jadhav; Mahesh Ravindra Khairnar; Rahul Gaybaro Naik
Journal:  J Prev Med Hyg       Date:  2021-01-14
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.