Measurement and correction of the effects of lag on contrast-detail test results in fluoroscopy.

N W Marshall1, C J Kotre.   

Abstract

Persistence of the video signal between TV frames, an effect also known as image lag, can lead to anomalously good contrast-detail test results for fluoroscopy systems. In this practical paper, a simple method is described which quantifies lag in fluoroscopy systems and corrects for its effect on threshold contrast. A digital framestore was used to acquire temporally contiguous fluoroscopy images. Correlation of the variance between an initial base TV frame and successive later frames was then measured via the correlation coefficient. Plotted against time, this function defines a time constant which characterizes the rate at which the initial variance pattern is replaced by incoming quantum noise. A survey of seven fluoroscopy units incorporating vacuum TV camera tubes found a mean time constant of 0.06 s. The relative change in contrast-detail performance was then measured as a function of applied digital frame averaging for two separate fluoroscopy units. A time constant was found for each frame averaging mode using the correlation of variance between frames. These measurements were used to derive a function which corrects contrast-detail results obtained for a unit with a measured nominal time constant to the typical vacuum camera tube time constant of 0.06 s. The correction is shown to significantly reduce the spread of contrast-detail results obtained over a range of temporal filtration settings.

Mesh:

Year:  2002        PMID: 11936180

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Solid-state fluoroscopic imager for high-resolution angiography: physical characteristics of an 8 cm x 8 cm experimental prototype.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan; Steven K Onishi
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

2.  Effect of detector lag on CT noise power spectra.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

3.  Novel method to determine recursive filtration and noise reduction in fluoroscopic imaging - a comparison of four different vendors.

Authors:  Bente Konst; Jacob Nøtthellen; Stine Nalum Naess; Magnus Båth
Journal:  J Appl Clin Med Phys       Date:  2020-12-14       Impact factor: 2.102

4.  Fluoroscopic Analysis of Tibial Translation in Anterior Cruciate Ligament Injured Knees With and Without Bracing During Forward Lunge.

Authors:  Maryam Jalali; Farzam Farahmand; Seyed Mohammad Ebrahim Mousavi; Seyed Ali Golestanha; Tahmineh Rezaeian; Shahram Shirvani Broujeni; Mehdi Rahgozar; Fateme Esfandiarpour
Journal:  Iran J Radiol       Date:  2015-07-22       Impact factor: 0.212

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.