[The use of a thyroid collar for intraoral radiography].

K Schmidt1, X L Velders, F C van Ginkel, P F van der Stelt.   

Abstract

OBJECTIVE: To determine whether a thyroid collar is a reasonable measure to reduce patient exposure from intraoral radiography (cost benefit analysis).
DESIGN: In the thyroid gland of a Rando phantom dose measurements were carried out to determine the effect of a thyroid collar during intraoral radiography.
SETTING: Department of Oral Radiology at ACTA, Amsterdam.
METHODS: Dose measurements were carried out using LTDs. The average absorbed dose to the thyroid gland with and without thyroid collar from intraoral radiography was compared using an analysis of variance.
RESULTS: For periapical radiographs the equivalent dose to the thyroid gland was significantly lower (p < 0.05) when a thyroid collar was used. For bitewing radiography there were no significant effects of the thyroid collar (p > 0.05). The cost benefit analysis showed that it takes more than 40 years before the benefits of a thyroid collar exceed the costs.
CONCLUSION: Collective use of thyroid collars therefore does not seem to be a reasonable measure to optimize radiological protection during intraoral radiography.

Entities:  

Mesh:

Year:  1998        PMID: 11928141

Source DB:  PubMed          Journal:  Ned Tijdschr Tandheelkd        ISSN: 0028-2200


  3 in total

1.  The value of thyroid shielding in intraoral radiography.

Authors:  Reinier C Hoogeveen; Bart Hazenoot; Gerard C H Sanderink; W Erwin R Berkhout
Journal:  Dentomaxillofac Radiol       Date:  2016-03-23       Impact factor: 2.419

2.  The use of a thyroid shield for intraoral anterior oblique occlusal views-a risk-based approach.

Authors:  Mark Worrall; Alison Menhinick; Donald J Thomson
Journal:  Dentomaxillofac Radiol       Date:  2017-10-27       Impact factor: 2.419

3.  Thyroid shields and neck exposures in cephalometric radiography.

Authors:  Philippe Hujoel; Lars Hollender; Anne-Marie Bollen; John D Young; Joana Cunha-Cruz; Molly McGee; Alex Grosso
Journal:  BMC Med Imaging       Date:  2006-06-13       Impact factor: 1.930

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.