OBJECTIVE: To evaluate whether coating AO/ASIF screws with osteoconductive materials can improve bone-screw fixation.
DESIGN: Ninety-six AO/ASIF 4.5-millimeter cortical screws were divided into four paired groups and implanted in the femurs and tibiae of six sheep: Group A = standard stainless steel screws; Group B = stainless steel screws coated with highly crystalline hydroxyapatite; Group C = stainless steel screws coated with low crystalline hydroxyapatite; Group D = titanium screws coated with titanium. The screws were implanted according to the standard AO technique to an insertion torque of 2,000 Newton-millimeters. Sheep were killed at one, three, and twelve months after surgery. MAIN OUTCOME MEASURES: Extraction torque was measured on six screws from each group selected at random at time of each euthanization. Morphologic analysis of the bone-screw interface was performed on the remaining screws.
RESULTS: At each euthanization the extraction torque of Group A was lower than that of the other groups (p < 0.0001). At three and twelve months the extraction torque of Group B was higher than that of Group D (p = 0.002). Morphologic results showed extensive bone-screw gap in Group A. Optimal osteointegration was observed in Groups B and C. Osteointegration of Group D was higher than that of Group A and lower than that of Groups B and C.
CONCLUSIONS: It was demonstrated that AO/ASIF screws coated with osteoconductive materials achieve optimal fixation strength, even in the early phase. This fixation strength was significantly higher than that of the standard screws.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11927807     DOI: 10.1097/00005131-200204000-00007

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  5 in total

1.  Microstructural analysis of implant-bone interface of hydroxyapatite-coated and uncoated Schanz screws.

Authors:  M Cimerman; A Cör; M Ceh; A Kristan; J Pizem; M Tonin
Journal:  J Mater Sci Mater Med       Date:  2005-07       Impact factor: 3.896

2.  A novel system for recording from single neurons in unrestrained animals.

Authors:  Helen Sherk; Elizabeth J Wilkinson
Journal:  J Neurosci Methods       Date:  2008-06-21       Impact factor: 2.390

Review 3.  From prevention of pin-tract infection to treatment of osteomyelitis during paediatric external fixation.

Authors:  Dimitri Ceroni; Catherine Grumetz; Odile Desvachez; Sophie Pusateri; Pierre Dunand; Eleftheria Samara
Journal:  J Child Orthop       Date:  2016-11-15       Impact factor: 1.548

4.  Screw stripping and its prevention in the hexagonal socket of 3.5-mm titanium locking screws.

Authors:  Hyo-Jin Lee; Young Uk Park; Sung Jae Kim; Hyong Nyun Kim
Journal:  Sci Rep       Date:  2021-10-29       Impact factor: 4.379

5.  A Simple Technique for Removal of Interlocking Screw from an Intramedullary Nail: A Case Report.

Authors:  Sumanth Madhusudan Prabhakar; Joshua Decruz; Wee Liang Hao James; Remesh Kunnasegaran
Journal:  J Orthop Case Rep       Date:  2021
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.