Abstract
A method for the determination of sulfonamides and trimethoprim in the complex matrix liquid manure has been developed using reversed-phase liquid chromatography and atmospheric pressure chemical ionization (APCI) tandem mass spectrometry. A comparison was made between electrospray and atmospheric pressure chemical ionization. APCI proved to be more robust and less sensitive to matrix effects. High-performance liquid chromatographic (HPLC) separation of the analytes was achieved in less than 7 min. The compounds were extracted with ethyl acetate and the extracts were cleaned up by solid-phase extraction on an aminopropyl column. Recoveries were not dependent on the concentration level. The mean recoveries were as follows: trimethoprim 79.0%, sulfadiazine 80.5%, N(4)-acetylsulfadiazine 91.0%, sulfamerazine 78.6%, sulfadimidine 77.2% and sulfamethoxazole 82.8%. Linearity was established over a concentration range of 5 to 5000 microg/kg with correlation coefficients greater than 0.99. The method had a limit of quantitation (LOQ) of 5 microg/kg manure. Copyright 2002 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2002 PMID: 11921244 DOI: 10.1002/rcm.624
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419