Literature DB >> 11914011

Bacterial colonisation of Doppler probes on vascular surgical wards.

S M G Kibria1, K G Kerr, J Dave, M J Gough, S Homer-Vanniasinkam, A I D Mavor.   

Abstract

AIM: hospital acquired infections cost the NHS 1 pound sterling billion each year and medical equipment may act both as source and vector of nosocomial infection. This study examined bacterial contamination of Doppler ultrasound probes (USP) in routine use on vascular surgical wards in six hospitals and the knowledge of staff about the potential for cross infection from contaminated probes.
METHODS: probe head impressions and swab cultures of probe holders were plated on mannitol salt agar before and after cleaning with a paper towel. Putative S. aureus isolates were identified to species level and susceptibility to selected antimicrobials tested. Concurrently, junior medical staff were surveyed about probe cleaning protocols.
RESULTS: methicillin susceptible S. aureus was isolated from 2/21 (10%) with near confluent bacterial growth from six others (28%). The latter may have obscured low numbers of S. aureus. Further since swabs were plated without prior enrichment culture, it is likely that contamination with S. aureus might have been underestimated. No positive cultures were obtained after wiping the USP with a paper towel. 22/23 (95%) junior doctors failed to clean the USP prior to use.
CONCLUSION: USP contamination with pathogenic bacteria occurs under "in-use" conditions and junior medical staff are unaware of simple measures to prevent this. Strict guidelines for USP cleaning between patient use should, therefore, be adopted particularly when monitoring postoperative graft patency. Copyright 2002 Harcourt Publishers Limited.

Entities:  

Mesh:

Year:  2002        PMID: 11914011     DOI: 10.1053/ejvs.2001.1552

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  4 in total

1.  Efficacy of soaking in 70% isopropyl alcohol on aerobic bacterial decontamination of surgical instruments and gloves for serial mouse laparotomies.

Authors:  Jessica N Keen; MaryKay Austin; Li-Shan Huang; Susan Messing; Jeffrey D Wyatt
Journal:  J Am Assoc Lab Anim Sci       Date:  2010-11       Impact factor: 1.232

2.  Contamination and decontamination of Doppler probes.

Authors:  Emma J Whitehead; John F Thompson; David R Lewis
Journal:  Ann R Coll Surg Engl       Date:  2006-09       Impact factor: 1.891

Review 3.  Clinical review: Bedside lung ultrasound in critical care practice.

Authors:  Bélaïd Bouhemad; Mao Zhang; Qin Lu; Jean-Jacques Rouby
Journal:  Crit Care       Date:  2007       Impact factor: 9.097

4.  The importance of infection prevention and control in medical ultrasound.

Authors:  Lia Moshkanbaryans; Craig Meyers; Andrew Ngu; Jon Burdach
Journal:  Australas J Ultrasound Med       Date:  2015-12-31
  4 in total