Literature DB >> 11914003

Removal of knotted intravascular devices. Case report and review of the literature.

I D Karanikas1, A Polychronidis, A Vrachatis, D P Arvanitis, C E Simopoulos, M K Lazarides.   

Abstract

OBJECTIVE: to review published reports on knotted intravascular devices/catheters.
METHOD: report of two cases and systematic review of the literature.
RESULTS: a total of 113 reported cases of knotted intravascular devices/catheters were located. Pulmonary artery catheters (Swan-Ganz) were responsible for more than two thirds of the total reported intravascular knots. In 62% (70/113) of the cases withdrawal of the knotted catheters was achieved successfully with different interventional radiological techniques, avoiding the need for surgical exploration. In 32% (36/113) of the patients surgical removal was favoured. Capture with one of the interventional techniques and pulling down the knot into an easily accessible vein to be removed through an open venotomy, was the most common surgical procedure. However, in five cases, an open cardiotomy was required. In seven cases the patient's condition was critical and precluded any surgical procedure, so the knotted catheter was left in situ. The mortality of this event was 8% (9/113).
CONCLUSIONS: interventional radiological techniques have largely replaced open surgical removal. Knotted catheters may need to be surgically removed when (a) the knot is large in size with many loops, or (b) intracardiac fixing of the knot is encountered. Copyright 2002 Harcourt Publishers Limited.

Entities:  

Mesh:

Year:  2002        PMID: 11914003     DOI: 10.1053/ejvs.2001.1591

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  15 in total

1.  Knotting of a pulmonary artery catheter in the superior vena cava: surgical removal and a word of caution.

Authors:  G P Georghiou; B A Vidne; E Raanani
Journal:  Heart       Date:  2004-05       Impact factor: 5.994

2.  [Knotted Seldinger wire].

Authors:  A Fiala; B Glodny; I H Lorenz
Journal:  Anaesthesist       Date:  2015-12-11       Impact factor: 1.041

3.  Transfemoral removal of a knotted Swan-Ganz catheter.

Authors:  John C Papakostas; Lazaros-Stavros N Papadopoulos; Helen M Arnaoutoglou; Angela Karahaliou; Miltiadis I Matsagas
Journal:  Can J Surg       Date:  2008-10       Impact factor: 2.089

4.  Transfemoral Technique to Untwist a Knotted Catheter in the Brachial Artery: A Case Report.

Authors:  Nathalie Noirclerc; Paul Alexandru Simion; Abdelkader Bakhti
Journal:  Maedica (Bucur)       Date:  2020-12

5.  Transfemoral contralateral technique to retrieve knotted coronary artery catheter using Amplatz Goose Neck snare catheter.

Authors:  Ihsan M Rafie; Girish Viswanathan; William J Penny
Journal:  BMJ Case Rep       Date:  2010-11-19

6.  Coil Knotting during Endovascular Coil Embolization for Ruptured MCA Aneurysm. A Case Report.

Authors:  S C Kwon; I U Lyo; S H Shin; J B Park; Y Kim
Journal:  Interv Neuroradiol       Date:  2008-10-08       Impact factor: 1.610

7.  How is that? Knotting of a peripherally inserted central venous catheter.

Authors:  Gnanamuthu Birla Roy; Ajit A Cheriyan; Manbha L Rymbai
Journal:  Indian J Anaesth       Date:  2010-05

8.  Percutaneous retrieval of centrally embolized fragments of central venous access devices or knotted Swan-Ganz catheters. Clinical report of 14 retrievals with detailed angiographic analysis and review of procedural aspects.

Authors:  Łukasz Kalińczuk; Zbigniew Chmielak; Artur Dębski; Cezary Kępka; Piotr N Rudziński; Sebastian Bujak; Mirosław Skwarek; Andrzej Kurowski; Zofia Dzielińska; Marcin Demkow
Journal:  Postepy Kardiol Interwencyjnej       Date:  2016-05-11       Impact factor: 1.426

9.  Jugular Venous Catheterization: A Case of Knotting.

Authors:  E Erkılıç; M Aksoy; T Gümüş; S Sarı; E Kesimci
Journal:  Case Rep Med       Date:  2015-11-08

10.  Successful percutaneous retrieval of unusually knotted temporary pacemaker lead.

Authors:  Jaywant M Nawale; Sandip N Patil; Digvijay D Nalawade; Nikhil A Borikar; Bhushan S Sonawane; Ajay S Chaurasia
Journal:  Indian Pacing Electrophysiol J       Date:  2018-03-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.