Literature DB >> 11901438

Transvenous transjugular renal core biopsy with a redesigned biopsy set including a blunt-tipped needle.

Constantinos T Sofocleous1, Philip Bahramipour, Christopher Mele, Clay R Hinrichs, Alison Barone, Hani Abujudeh.   

Abstract

A novel 19-gauge, blunt-tipped, side cutting single throw, 70-cm long transjugular needle, specifically designed for transvenous kidney biopsy, was used in seven patients with high risk for bleeding. A mean of 4 device-passes (3-6) per patient resulted in a satisfactory specimen for pathological diagnosis. Immediate post-biopsy nonenhanced CT was performed to evaluate for bleeding at the biopsy site. All patients were observed for 2 hr after the procedure. No clinically significant immediate or late complication was noted.

Entities:  

Mesh:

Year:  2002        PMID: 11901438     DOI: 10.1007/s00270-001-0093-8

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  3 in total

Review 1.  Complications of transjugular biopsies.

Authors:  Rakesh Navuluri; Osman Ahmed
Journal:  Semin Intervent Radiol       Date:  2015-03       Impact factor: 1.513

2.  Technical safety and efficacy of a blunt-tip microwave ablation electrode for CT-guided ablation of pulmonary ground-glass opacity nodules.

Authors:  Jiachang Chi; Zhi Wang; Min Ding; Hao Hu; Bo Zhai
Journal:  Eur Radiol       Date:  2021-04-02       Impact factor: 5.315

3.  Safety and effectiveness of transjugular renal biopsy: A single center study.

Authors:  K R Rathod; B A Popat; A Pandey; T E Jamale; N K Hase; H L Deshmukh
Journal:  Indian J Nephrol       Date:  2017 Mar-Apr
  3 in total