Literature DB >> 11888959

Adequately washed bronchoscope does not induce false-positive amplification tests on bronchial aspirates in the diagnosis of pulmonary tuberculosis.

Tae Sun Shim1, Hyun Sook Chi, Sang Do Lee, Younsuck Koh, Woo Sung Kim, Dong Soon Kim, Won Dong Kim.   

Abstract

STUDY
OBJECTIVES: To investigate the clinical usefulness of amplification (COBAS AMPLICOR; Roche Diagnostics Systems; Branchburg, NJ) on bronchoscopic aspirate specimens in the diagnosis of pulmonary tuberculosis, with particular regard to the possibility of false-positive results in subsequent specimens due to residual Mycobacterium tuberculosis DNA. DESIGN AND
SETTING: A prospective clinical study at a tertiary referral medical center. PARTICIPANTS AND METHODS: Four hundred fiberoptic bronchoscopic procedures were performed, using seven bronchoscopes on 335 consecutive patients, for therapeutic or diagnostic purposes. Serial bronchial aspirates were collected and tested for M tuberculosis, using COBAS AMPLICOR (CA). Bronchoscopes were cleaned and disinfected automatically, between patient use, by the same endoscope washer. The name of each bronchoscope and the sequence of its use were recorded, together with the sequence of washing. The CA results were compared with the bacteriologic and histologic results for M tuberculosis infection. When there was a suspicion of contamination, outward polymerase chain reaction analysis was performed.
RESULTS: Of 392 specimens (332 subjects), excluding the 8 specimens (4 subjects) in which bacteriologic and histologic analyses were omitted, a smear-positive result for acid-fast bacilli (AFB), culture-positive or biopsy-positive results, and CA-positive results were obtained in 16, 49, and 32 specimens, respectively. In AFB smear-positive subjects, the sensitivity, specificity, positive predictive values (PPVs), and negative predictive values (NPVs) were 92%, 67%, 92%, and 67%, respectively. In AFB smear-negative subjects, the sensitivity, specificity, PPV, and NPV values were 38%, 99%, 74%, and 94%, respectively. The CA test was more sensitive than the AFB smears for the diagnosis of pulmonary tuberculosis (53% vs 27%, respectively; p < 0.05). False-positive CA results were seen in only six specimens. Three of these six subjects received a diagnosis of pulmonary tuberculosis on clinical and radiologic grounds, and none of the six results seemed to be associated with bronchoscopic cross-contamination.
CONCLUSIONS: Adequately cleaned and disinfected bronchoscopes did not cause false-positive amplification test results for M tuberculosis on bronchial aspirates by cross-contamination. Furthermore, sensitivity was greater with the CA tests. Therefore, CA tests on bronchial aspirates seem to be useful in the diagnosis of pulmonary tuberculosis.

Entities:  

Mesh:

Year:  2002        PMID: 11888959     DOI: 10.1378/chest.121.3.774

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  2 in total

1.  Comparison of sterilization of reusable endoscopic biopsy forceps by autoclaving and ethylene oxide gas.

Authors:  Jai Hoon Yoon; Byung Chul Yoon; Hang Lak Lee; Jun Kyu Lee; Yong-Tae Kim; Dong Ho Lee; Il Ju Choi; Don Haeng Lee; Dong Hee Kim
Journal:  Dig Dis Sci       Date:  2011-09-09       Impact factor: 3.199

2.  Commercial nucleic-acid amplification tests for diagnosis of pulmonary tuberculosis in respiratory specimens: meta-analysis and meta-regression.

Authors:  Daphne I Ling; Laura L Flores; Lee W Riley; Madhukar Pai
Journal:  PLoS One       Date:  2008-02-06       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.