Literature DB >> 11856564

A three-dimensional multi-electrode array for multi-site stimulation and recording in acute brain slices.

Marc Olivier Heuschkel1, Michael Fejtl, Mario Raggenbass, Daniel Bertrand, Philippe Renaud.   

Abstract

Several multi-electrode array devices integrating planar metal electrodes were designed in the past 30 years for extracellular stimulation and recording from cultured neuronal cells and organotypic brain slices. However, these devices are not well suited for recordings from acute brain slice preparations due to a dead cell layer at the tissue slice border that appears during the cutting procedure. To overcome this problem, we propose the use of protruding 3D electrodes, i.e. tip-shaped electrodes, allowing tissue penetration in order to get closer to living neurons in the tissue slice. In this paper, we describe the design and fabrication of planar and 3D protruding multi-electrode arrays. The electrical differences between planar and 3D protruding electrode configuration were simulated and verified experimentally. Finally, a comparison between the planar and 3D protruding electrode configuration was realized by stimulation and recording from acute rat hippocampus slices. The results show that larger signal amplitudes in the millivolt range can be obtained with the 3D electrode devices. Spikes corresponding to single cell activity could be monitored in the hippocampus CA3 and CA1 region using 3D electrodes.

Entities:  

Mesh:

Year:  2002        PMID: 11856564     DOI: 10.1016/s0165-0270(01)00514-3

Source DB:  PubMed          Journal:  J Neurosci Methods        ISSN: 0165-0270            Impact factor:   2.390


  51 in total

1.  Understanding and improving photo-control of ion channels in nociceptors with azobenzene photo-switches.

Authors:  Alexandre Mourot; Christian Herold; Michael A Kienzler; Richard H Kramer
Journal:  Br J Pharmacol       Date:  2017-07-27       Impact factor: 8.739

2.  A lithographically-patterned, elastic multi-electrode array for surface stimulation of the spinal cord.

Authors:  Kathleen W Meacham; Richard J Giuly; Liang Guo; Shawn Hochman; Stephen P DeWeerth
Journal:  Biomed Microdevices       Date:  2008-04       Impact factor: 2.838

3.  Multilayer PDMS microfluidic chamber for controlling brain slice microenvironment.

Authors:  A J Blake; T M Pearce; N S Rao; S M Johnson; J C Williams
Journal:  Lab Chip       Date:  2007-05-24       Impact factor: 6.799

4.  Modeling the nonlinear properties of the in vitro hippocampal perforant path-dentate system using multielectrode array technology.

Authors:  Angelika Dimoka; Spiros H Courellis; Ghassan I Gholmieh; Vasilis Z Marmarelis; Theodore W Berger
Journal:  IEEE Trans Biomed Eng       Date:  2008-02       Impact factor: 4.538

5.  Causal pattern recovery from neural spike train data using the Snap Shot Score.

Authors:  Christoph Echtermeyer; Tom V Smulders; V Anne Smith
Journal:  J Comput Neurosci       Date:  2009-07-31       Impact factor: 1.621

6.  Three-dimensional micro-electrode array for recording dissociated neuronal cultures.

Authors:  Katherine Musick; David Khatami; Bruce C Wheeler
Journal:  Lab Chip       Date:  2009-04-08       Impact factor: 6.799

7.  Monitoring hippocampus electrical activity in vitro on an elastically deformable microelectrode array.

Authors:  Zhe Yu; Oliver Graudejus; Candice Tsay; Stéphanie P Lacour; Sigurd Wagner; Barclay Morrison
Journal:  J Neurotrauma       Date:  2009-07       Impact factor: 5.269

8.  Perfused drop microfluidic device for brain slice culture-based drug discovery.

Authors:  Jing Liu; Liping Pan; Xuanhong Cheng; Yevgeny Berdichevsky
Journal:  Biomed Microdevices       Date:  2016-06       Impact factor: 2.838

9.  Experimental Investigation on Spontaneously Active Hippocampal Cultures Recorded by Means of High-Density MEAs: Analysis of the Spatial Resolution Effects.

Authors:  Alessandro Maccione; Mauro Gandolfo; Mariateresa Tedesco; Thierry Nieus; Kilian Imfeld; Sergio Martinoia; Luca Berdondini
Journal:  Front Neuroeng       Date:  2010-05-10

10.  High efficient electrical stimulation of hippocampal slices with vertically aligned carbon nanofiber microbrush array.

Authors:  Edward D de Asis; T D Barbara Nguyen-Vu; Prabhu U Arumugam; Hua Chen; Alan M Cassell; Russell J Andrews; Cary Y Yang; Jun Li
Journal:  Biomed Microdevices       Date:  2009-08       Impact factor: 2.838

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.